Suppr超能文献

在 D 波记录期间 BIS 监测的 EEG 迹线上的刺激伪影。

Stimulation artefact on EEG trace with BIS monitoring during D-wave recording.

机构信息

Department of Anaesthesia, Sakra world hospital, 560103, Bellandur, Bengaluru, Karnataka, India.

Department of Neurosciences, Sakra world hospital, 560103, Bellandur, Bengaluru, Karnataka, India.

出版信息

J Clin Monit Comput. 2023 Jun;37(3):929-931. doi: 10.1007/s10877-022-00951-w. Epub 2022 Dec 12.

Abstract

The D-wave reflects the corticospinal fibre potentials and is frequently recorded intraoperatively for intramedullary spinal tumours to ensure maximum safe resection. A 38-year-old male had an intramedullary spinal cord tumour surgically removed while being monitored with D-wave from a single distal electrode technique. Total intravenous anaesthesia (TIVA) with propofol (without a muscle relaxant and with intermittent intravenous fentanyl boluses) was used for the maintenance of anaesthesia guided by processed electroencephalography (EEG) using a bispectral index (BIS) monitor. Regular spike artefacts were observed in the EEG signal recorded by the BIS® monitor during the application of the single-pulse transcranial electrical stimulus and were used as a visual indicator of stimulus delivery. Finally, we propose a novel method of confirming stimulus delivery during D-wave recording based on stimulation artefacts in the EEG signal recorded by the BIS® monitor.

摘要

D 波反映皮质脊髓纤维的电位,常用于监测髓内脊髓肿瘤,以确保最大限度的安全切除。一名 38 岁男性在接受全身麻醉下(无肌松剂,间断静脉注射芬太尼),采用脑电双频指数(BIS)监测下的丙泊酚全凭静脉麻醉(TIVA),使用单脉冲经颅电刺激,并根据处理后的脑电图(EEG)使用双频谱指数(BIS)监测仪进行麻醉诱导。在应用单脉冲经颅电刺激时,BIS®监测仪记录的 EEG 信号中观察到常规的尖波伪迹,并将其用作刺激传递的视觉指标。最后,我们提出了一种基于 BIS®监测仪记录的 EEG 信号中刺激伪迹来确认 D 波记录中刺激传递的新方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验