Medical Physics Department, Veneto Institute of Oncology IOV-IRCCS, Padova, Italy.
Dipartimento di Fisica e Astronomia "G. Galilei", Università degli Studi di Padova, Padova, Italy.
J Appl Clin Med Phys. 2023 Jan;24(1):e13781. doi: 10.1002/acm2.13781. Epub 2022 Dec 15.
An unnecessary amount of complexity in radiotherapy plans affects the efficiency of the treatments, increasing the uncertainty of dose deposition and its susceptibility to anatomical changes or setup errors. To date, tools for quantitatively assessing the complexity of tomotherapy plans are still limited. In this study, new metrics were developed to characterize different aspects of helical tomotherapy (HT) plans, and their actual effectiveness was investigated.
The complexity of 464 HT plans delivered on a Radixact platform was evaluated. A new set of metrics was devised to assess beam geometry, leaf opening time (LOT) variability, and modulation over space and time. Sixty-five complexity metrics were extracted from the dataset using the newly in-house developed software library TCoMX: 29 metrics already proposed in the literature and 36 newly developed metrics. Their reciprocal relation is discussed. Their effectiveness was evaluated through correlation analyses with patient-specific quality assurance (PSQA) results.
An inverse linear relation was found between the average number of closed leaves and the average number of MLC openings and closures as well as between the choice of the modulation factor and the discontinuity of the field, suggesting some intrinsic link between the LOT distribution and the geometrical complexity of the MLC openings. The newly proposed metrics were at least as correlated as the existing ones to the PSQA results. Metrics describing the geometrical complexity of the MLC openings showed the strongest connection to the PSQA results. Significant correlations were found between at least one of the new metrics and the γ index passing rate for six out of seven groups of plans considered.
The new metrics proposed were shown to be effective to characterize more comprehensively the complexity of HT plans. A software library for their automatic extraction is described and made available.
放疗计划中不必要的复杂性会影响治疗效率,增加剂量沉积的不确定性,并使其容易受到解剖结构变化或设置误差的影响。迄今为止,用于定量评估调强放疗计划复杂性的工具仍然有限。在这项研究中,开发了新的指标来描述螺旋断层放疗(HT)计划的不同方面,并研究了它们的实际效果。
评估了在 Radixact 平台上交付的 464 个 HT 计划的复杂性。设计了一组新的指标来评估射束几何形状、叶片开启时间(LOT)变化以及空间和时间上的调制。使用新开发的内部 TCoMX 软件库从数据集提取了 65 个复杂性指标:29 个文献中已提出的指标和 36 个新开发的指标。讨论了它们的倒数关系。通过与患者特定质量保证(PSQA)结果的相关分析评估了它们的有效性。
发现平均闭合叶片数与平均 MLC 开启和关闭数之间以及调制因子的选择与场的不连续性之间呈线性反比关系,这表明 LOT 分布与 MLC 开口的几何复杂性之间存在某种内在联系。新提出的指标与 PSQA 结果的相关性至少与现有的指标一样高。描述 MLC 开口几何复杂性的新指标与 PSQA 结果的相关性最强。对于所考虑的七个计划组中的六个组,至少有一个新指标与 γ 指数通过率之间存在显著相关性。
所提出的新指标被证明可以有效地更全面地描述 HT 计划的复杂性。描述了一个用于自动提取它们的软件库,并提供了该库。