Lee Jason K Y, Cradic Kendall, Singh Ravinder J, Jones JoAnna, Li Jieli
Department of Clinical Laboratory, University Hospital, The Ohio State University Wexner Medical Center, Columbus, OH, United States.
OhioHealth Laboratory Services, Department of Pathology, OhioHealth, Columbus, OH, United States; CORPath Pathology Services, LLC, Columbus, OH, United States.
Clin Chim Acta. 2023 Jan 15;539:130-133. doi: 10.1016/j.cca.2022.11.034. Epub 2022 Dec 14.
Inconsistent Insulin-like Growth Factor 1 (IGF-1) measurements among different platforms have been observed. In this study, we compared the IGF-1 assay on four different platforms.
A total of 110 serum specimens were analyzed in this comparison study. IGF-1 was measured on the three different chemiluminescent automated immunoassay of Siemens Immulite 2000 XPi, DiaSorin Liaison XL, IDS iSYS and LC-MS/MS method. Results were compared with Weighted Deming regression. Bias was evaluated using the Bland-Altman method.
Weighted Deming regression analysis showed approximately 36 % negative variation on Immulite, compared to Liaison (Immulite = 0.64 * DiaSorin + 2.95, r = 0.95); 8 % negative variation on iSYS, compared to Liaison (iSYS = 0.92 * DiaSorin + 0.51, r = 0.97); 17 % negative variation on LC-MS/MS, compared to Liaison (LC-MS/MS = 0.83 * DiaSorin-11.23, r = 0.93); 34 % positive variation on LC-MS/MS compared to Immulite (LC-MS/MS = 1.34 * Immulite-21.97, r = 0.96); 81 % positive variation on IDS iSYS compared to Immulite (IDS iSYS = 1.81 * Immulite-117.65, r = 0.83). The Bland-Altman plot showed a significant negative variation of Immulite versus DiaSorin and positive variation of IDS iSYS versus Immulite. Overall agreement between different platforms was poor, which reflected systematic difference. The variation between platforms increased as IGF-1 values increased.
There are wide variations between different platforms for IGF-1 measurement. The lack of standardization in IGF-1 measurement creates a challenge for clinicians to monitor IGF-1 and treat patients with pituitary disorders, when switching from one platform to another. The potential impact of the variations in IGF-1 measurement between different platforms should be taken into consideration when managing patients.
不同检测平台间胰岛素样生长因子1(IGF-1)检测结果不一致。本研究比较了四种不同平台上的IGF-1检测方法。
本比较研究共分析了110份血清标本。采用西门子Immulite 2000 XPi、索灵Liaison XL、IDS iSYS的三种不同化学发光自动化免疫分析方法及液相色谱-串联质谱法(LC-MS/MS)检测IGF-1。结果采用加权德明回归法进行比较。采用Bland-Altman法评估偏倚。
加权德明回归分析显示,与Liaison相比,Immulite约有36%的负偏差(Immulite = 0.64 * 索灵 + 2.95,r = 0.95);与Liaison相比,iSYS有8%的负偏差(iSYS = 0.92 * 索灵 + 0.51,r = 0.97);与Liaison相比,LC-MS/MS有17%的负偏差(LC-MS/MS = 0.83 * 索灵 - 11.23,r = 0.93);与Immulite相比,LC-MS/MS有34%的正偏差(LC-MS/MS = 1.34 * Immulite - 21.97,r = 0.96);与Immulite相比,IDS iSYS有81%的正偏差(IDS iSYS = 1.81 * Immulite - 117.65,r = 0.83)。Bland-Altman图显示Immulite与索灵相比有显著负偏差,IDS iSYS与Immulite相比有正偏差。不同平台间的总体一致性较差,这反映出存在系统差异。随着IGF-1值升高,各平台间的差异增大。
不同平台间IGF-1检测结果差异很大。IGF-1检测缺乏标准化给临床医生在监测IGF-1及治疗垂体疾病患者时从一个平台转换至另一个平台带来了挑战。在管理患者时,应考虑不同平台间IGF-1检测差异的潜在影响。