Reck R, Störkel S, Meyer A
Hals-Nasen-Ohrenklinik, Universität Mainz.
Laryngol Rhinol Otol (Stuttg). 1987 Jul;66(7):373-6.
An 8-year follow-up of Ceravital middle ear prostheses showed an extremely good tolerance in the middle ear space. Incompatibility phenomena were not observed. Inflammatory reactions were neither caused nor supported by the implants. The implantation was always performed without interposition of cartilage. Extrusions were not observed. Long-lasting inflammatory processes destroy implants in the same way as they affect ossicles. Transient inflammatory periods (e.g. in the early postoperative period, POM) do not lead to any remarkable changes of the implant surface. In healthy middle ears, the hearing results are stable.
对Ceravital中耳假体进行的8年随访显示,其在中耳腔的耐受性极佳。未观察到不相容现象。植入物既未引发也未助长炎症反应。植入手术始终在不插入软骨的情况下进行。未观察到假体脱出。长期的炎症过程会像影响听小骨一样破坏植入物。短暂的炎症期(如术后早期)不会导致植入物表面出现任何明显变化。在健康的中耳中,听力结果稳定。