Suppr超能文献

参考电极在电阻抗断层成像中的影响。

The influence of reference electrode in electrical impedance tomography.

作者信息

Lin Zhimin, Huang Weixiang, Gao Zhijun, Yang Lin, Li Yimin, Lu Yu, Dai Meng, Fu Feng, Sang Ling, Zhao Zhanqi

机构信息

State Key Lab of Respiratory Diseases, Guangzhou Institute of Respiratory Health, Guangzhou Medical University, The First Affiliated Hospital of Guangzhou Medical University, Department of Critical Care Medicine, Guangzhou, China.

Department of Aerospace Medicine, Air Force Medical University, Xi'an, China.

出版信息

Heliyon. 2022 Dec 21;8(12):e12454. doi: 10.1016/j.heliyon.2022.e12454. eCollection 2022 Dec.

Abstract

BACKGROUND

Some electrical impedance tomography (EIT) devices equip reference electrodes. In practice, it is not uncommon to observe high contact impedance for the reference electrode. The influence of bad contact reference electrode on data quality is unknown. The study aimed to investigate the influence of reference electrode on EIT image reconstruction.

METHODS

Thirty lung healthy volunteers were prospectively examined with EIT. The subjects were spontaneously breathing in supine position. Three scenarios were constructed: 1. Normal measurement; 2. Reference electrode disconnected without recalibration; 3. Reference electrode disconnected, and the measurement restarted after recalibration of the system. EIT-based parameters measuring spatial and temporal ventilation distributions were calculated and compared. A so-call deviation score was calculated to assess the differences in EIT parameters between scenarios 2 and 1, between 3 and 1.

RESULTS

The absolute differences for all parameters were significantly higher than zero (p < 0.01 for all parameters and scenarios). Deviation score for scenario 2 was 4.5 ± 3.5. Four subjects had a deviation score of 0 in scenario 2 and five subjects had a score of 1. The deviation in scenario 3 was higher (6.1 ± 3.1). No subjects had a score of 0 and only two subjects had a score of 1.

CONCLUSIONS

For EIT systems that equips with reference electrode, it is important to ensure the proper contact and functionality of the reference electrode. The EIT data quality would remain unchanged in only a small portion of subjects.

摘要

背景

一些电阻抗断层成像(EIT)设备配备了参考电极。在实际操作中,参考电极出现高接触阻抗的情况并不少见。接触不良的参考电极对数据质量的影响尚不清楚。本研究旨在探讨参考电极对EIT图像重建的影响。

方法

对30名肺部健康志愿者进行EIT前瞻性检查。受试者仰卧位自主呼吸。构建了三种情况:1. 正常测量;2. 参考电极断开且未重新校准;3. 参考电极断开,系统重新校准后重新开始测量。计算并比较基于EIT测量空间和时间通气分布的参数。计算所谓的偏差分数,以评估情况2与情况1之间、情况3与情况1之间EIT参数的差异。

结果

所有参数的绝对差异均显著高于零(所有参数和情况的p<0.01)。情况2的偏差分数为4.5±3.5。4名受试者在情况2中的偏差分数为0,5名受试者的分数为1。情况3中的偏差更高(6.1±3.1)。没有受试者的分数为0,只有2名受试者的分数为1。

结论

对于配备参考电极的EIT系统,确保参考电极的正确接触和功能非常重要。只有一小部分受试者的EIT数据质量会保持不变。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/eea4/9800185/8cfc111cb47b/gr1.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验