输液管路长度和注射器大小对输液系统顺应性的影响:使用低流速注射器型输液泵的实验研究。

Effects of infusion tubing line lengths and syringe sizes on infusion system compliance: an experimental study using a syringe-type infusion pump at low flow rate.

机构信息

Department of Anesthesiology and Pain Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, 81 Irwon-ro, Gangnam-gu, Seoul, 06351, Korea.

出版信息

J Clin Monit Comput. 2023 Oct;37(5):1379-1386. doi: 10.1007/s10877-023-00973-y. Epub 2023 Jan 17.

Abstract

Ideally, the flow delivery of an infusion system is proportional only to the rate of mechanical actuation of the syringe pump plunger. However, in the real world, overall infusion system compliance may be affected by components such as an extension of tubing lines, or different sizes of syringes. With higher compliance, there may be greater chances of flow irregularity. In this experimental study, we investigated the effects of lengths of infusion lines and syringe sizes on the compliance of syringe pumps with low flow rate (2 ml h). In the first experiment, infusion system compliance was measured in various settings by occlusion release. As the infusion tubing length and size of the syringe increased, the time to reach each pressure was delayed and the infusion system compliance increased. The contributions to system compliance from syringes were significantly greater compared to those of extended infusion lines. In the occlusion alarm experiment, the occlusion alarm could be delayed by 69.76 ± 3.98 min for the 50-ml syringe with a 560 cm infusion line set-up. In conclusion, the compliance of a syringe pump system increases as the loaded syringe size, or the length of the infusion tubing increases. The occlusion alarm may be much delayed and not useful in highly compliant systems with respect to the potential occlusion of the infusion system, so more attention is required when using a highly compliant infusion system.

摘要

理想情况下,输注系统的流速仅与注射器泵柱塞的机械驱动速度成比例。然而,在现实世界中,整体输注系统顺应性可能会受到诸如延长的管线路径或不同大小的注射器等因素的影响。顺应性较高,可能会有更大的流速不规则的机会。在这项实验研究中,我们研究了输注管路长度和注射器大小对低流速(2ml/h)注射器泵顺应性的影响。在第一个实验中,通过释放阻塞来测量各种设置下的输注系统顺应性。随着输注管路长度和注射器尺寸的增加,达到每个压力的时间延迟,输注系统顺应性增加。与延长的输注管路相比,注射器对系统顺应性的贡献明显更大。在阻塞报警实验中,对于设置有 560cm 输注管路的 50ml 注射器,阻塞报警可延迟 69.76±3.98 分钟。总之,随着加载的注射器尺寸或输注管路长度的增加,注射器泵系统的顺应性增加。对于输注系统潜在阻塞的高度顺应性系统,阻塞报警可能会延迟很长时间,并且没有用处,因此在使用高度顺应性的输注系统时需要更加注意。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验