乳腺癌患者适形治疗计划近表面区域中治疗计划系统计算剂量与 Markus 电离室测量剂量的剂量一致性分析

Dose Agreement Analysis of Treatment Planning System-Calculated Doses and Markus Chamber-Measured Doses in the Near-Surface Region for Breast Cancer Patients' Conformal Treatment Plans.

作者信息

Gaur Garima, Banipal Raja Paramjeet Singh, Garg Pardeep, Gurjar Om Prakash, Kaur Gurpreet, Sachdeva Sheetal, Grover Romikant, Kang Manraj Singh, Dangwal Vinod Kumar

机构信息

Department of Radiation Oncology, Guru Gobind Singh Medical College and Hospital, Faridkot, Punjab, India.

Department of Radiation Oncology, Government Medical College, Patiala, Punjab, India.

出版信息

J Med Phys. 2022 Jul-Sep;47(3):256-261. doi: 10.4103/jmp.jmp_153_21. Epub 2022 Nov 8.

Abstract

BACKGROUND

Surface/skin dose measurement is one of the most challenging tasks for clinical dosimetry in radiotherapy and comparison with almost all the commercially available treatment planning systems (TPSs) brings a significant variation with the measured dose.

AIMS AND OBJECTIVES

In the current study, doses calculated from the TPS in the near-surface region for conformal plans (both three-dimensional conformal radiotherapy [3DCRT] and intensity-modulated radiotherapy [IMRT]) of 35 breast cancer patients were evaluated and compared with the doses measured with Markus chamber.

MATERIALS AND METHODS

The computed tomography (CT) images of a solid water slab phantom with a Markus chamber (at different depths ranging from 1 mm to 5 mm from the surface) were taken and imported into the TPS. All the conformal treatment plans made in TPS were executed on a linear accelerator and dose agreements between TPS calculated and chamber measured doses were analysed.

RESULTS

Results showed that this TPS underestimated the calculated doses in the superficial region by up to 26% and 21%, respectively, with respect to mean and maximum dose values obtained within the effective volume of the chamber used.

CONCLUSION

The uncertainty of doses in the superficial region should be kept in mind when evaluating treatment plans for superficial tumours in TPS.

摘要

背景

表面/皮肤剂量测量是放射治疗临床剂量测定中最具挑战性的任务之一,并且与几乎所有市售治疗计划系统(TPS)的比较都显示出与测量剂量存在显著差异。

目的

在本研究中,对35例乳腺癌患者适形计划(三维适形放疗[3DCRT]和调强放疗[IMRT])在近表面区域由TPS计算得出的剂量进行评估,并与用 Markus 电离室测量的剂量进行比较。

材料与方法

获取带有Markus电离室(位于距表面1毫米至5毫米的不同深度处)的固体水模体的计算机断层扫描(CT)图像,并导入TPS。在直线加速器上执行TPS中制定的所有适形治疗计划,并分析TPS计算剂量与电离室测量剂量之间的剂量一致性。

结果

结果表明,相对于在所使用电离室有效体积内获得的平均剂量值和最大剂量值,该TPS在浅表区域分别低估计算剂量达26%和21%。

结论

在评估TPS中浅表肿瘤的治疗计划时,应牢记浅表区域剂量的不确定性。