Biomed Phys Eng Express. 2023 Feb 3;9(2). doi: 10.1088/2057-1976/acb551.
Radiation of thoracic computed tomography (CT) involves the breast although it is not considered an organ of interest. According to the International Commission on Radiological Protection (ICRP) No. 103, the breast is an organ with a high level of sensitivity when interacting with x-rays, increasing the potential risk of breast cancer. Therefore, the radiation dose must be optimized while maintaining image quality. The dose optimization can be accomplished using a radiation shield. This study aims to determine the effect of silicone rubber (SR)-lead (Pb) in various thicknesses as an alternative protective material limiting dose and preserving the image quality of the breast in thoracic CT. SR-Pb was made from SR and Pb by a simple method. The SR-Pb had thicknesses of 3, 6, 9, and 12 mm. The breast dose was measured using a CT dose profiler on the surface of the breast phantom. The CT number and the noise level of the resulting image were determined quantitatively. The dose without the radiation shield was 5.4 mGy. The doses measured using shielding with thicknesses of 3, 6, 9, and 12 mm were 5.2, 4.5, 4.3, and 3.3 mGy, respectively. Radiation shielding with a thickness of 12 mm reduced breast surface dose by up to 38%. The CT numbers and noise levels for the left and right breast phantom images were almost the same as those without radiation shields indicating there were only slight artifacts in the image. Therefore, SR-Pb is considered a good shielding material which can be pplied in a clinical setting by placing it directly on the breast surface for dose optimization.
胸部计算机断层扫描(CT)的辐射虽然不被认为是感兴趣的器官,但会涉及到乳房。根据国际辐射防护委员会(ICRP)第 103 号报告,乳房是与 X 射线相互作用时具有高敏感度的器官,增加了患乳腺癌的潜在风险。因此,必须在保持图像质量的同时优化辐射剂量。辐射剂量的优化可以使用辐射屏蔽来实现。本研究旨在确定不同厚度的硅橡胶(SR)-铅(Pb)作为替代防护材料限制剂量并保持胸部 CT 中乳房图像质量的效果。SR-Pb 是通过简单的方法由 SR 和 Pb 制成的。SR-Pb 的厚度分别为 3、6、9 和 12 毫米。使用 CT 剂量分布仪在乳房体模表面测量乳房剂量。定量确定所得图像的 CT 数和噪声水平。无辐射屏蔽的剂量为 5.4 mGy。使用厚度为 3、6、9 和 12 毫米的屏蔽测量的剂量分别为 5.2、4.5、4.3 和 3.3 mGy。厚度为 12 毫米的辐射屏蔽可将乳房表面剂量降低高达 38%。左、右乳房体模图像的 CT 数和噪声水平与无辐射屏蔽时几乎相同,表明图像中只有轻微的伪影。因此,SR-Pb 被认为是一种良好的屏蔽材料,可以直接放置在乳房表面,通过优化剂量来应用于临床环境。