Balancing Radiation Dose Reduction and Image Quality in Chest Computed Tomography using Silicon Rubber-barium Sulfate Composite Shield.

作者信息

Keshtkar Mohammad, Yazdanifar Saeedeh

机构信息

Department of Medical Physics and Radiology, Infectious Diseases Research Centre, Faculty of Medicine, Gonabad University of Medical Sciences, Gonabad, Iran.

Student Research Committee, Gonabad University of Medical Sciences, Gonabad, Iran.

出版信息

J Med Signals Sens. 2025 Jul 10;15:20. doi: 10.4103/jmss.jmss_61_24. eCollection 2025.

Abstract

BACKGROUND

During chest CT examinations, the breasts are exposed to a significant amount of radiation, increasing the risk of radiation-induced cancers. The objective of this study is to develop and evaluate a novel silicon rubber-barium sulfate (BaSO4) composite breast shield for reducing radiation dose in chest computed tomography (CT) examinations while minimizing impact on image quality.

METHODS

Four breast shields were fabricated: one with 10% bismuth and three with 10%, 15%, and 20% BaSO4. Dose reduction was assessed using a thorax phantom and ionization chamber. Image quality effects were evaluated in the thorax phantom by measuring noise and CT number changes. The 10% barium shield was further tested on 22 patients undergoing chest CT.

RESULTS

The 10%, 15%, and 20% barium shields reduced breast dose by 36.8%, 38.6%, and 45.6%, respectively, while the 10% bismuth shield achieved a 63.1% reduction. However, the 10% barium shield had minimal impact on image quality, increasing lung noise by only 0.3 Hounsfield units (HU) and shifting CT numbers by 4.7 HU. In patient studies, 81.8% of scans showed no artifacts, with 18.2% showing slight artifacts.

CONCLUSION

The 10% BaSO4 shield effectively reduced breast dose while maintaining image quality, presenting a viable alternative to bismuth shielding for radiation protection in chest CT examinations.

摘要

背景

在胸部CT检查过程中,乳房会受到大量辐射,从而增加了辐射诱发癌症的风险。本研究的目的是开发并评估一种新型的硅橡胶 - 硫酸钡(BaSO4)复合乳房防护罩,以降低胸部计算机断层扫描(CT)检查中的辐射剂量,同时尽量减少对图像质量的影响。

方法

制作了四个乳房防护罩:一个含10%铋,另外三个分别含10%、15%和20%的硫酸钡。使用胸部体模和电离室评估剂量降低情况。通过测量噪声和CT值变化,在胸部体模中评估图像质量影响。对22名接受胸部CT检查的患者进一步测试了含10%钡的防护罩。

结果

含10%、15%和20%钡的防护罩分别使乳房剂量降低了36.8%、38.6%和45.6%,而含10%铋的防护罩剂量降低了63.1%。然而,含10%钡的防护罩对图像质量的影响最小,仅使肺部噪声增加0.3亨氏单位(HU),CT值偏移4.7 HU。在患者研究中,81.8%的扫描未显示伪影,18.2%显示轻微伪影。

结论

10%硫酸钡防护罩在有效降低乳房剂量的同时保持了图像质量,是胸部CT检查中用于辐射防护的铋屏蔽的可行替代方案。