四种根尖定位仪在有牙根吸收的前磨牙中的准确性:一项研究。

Accuracy of Four Apex Locators in Premolars with Root Resorption: An Study.

作者信息

López-Maekawa Astrid, García Rupay Carmen Rosa

机构信息

Endodontics Specialist, Master Program in Stomatology, School of Stomatology, Universidad Cientifica del Sur, Lima, Peru.

Endodontics Specialist, Master of Stomatology, School of Stomatology, Universidad Cientifica del Sur, Lima, Peru.

出版信息

Iran Endod J. 2022 Winter;17(1):20-26. doi: 10.22037/iej.v17i1.35498.

Abstract

INTRODUCTION

The aim of the present study was to compare the accuracy of the working length measurements of four foramen locators: Root ZX mini, Raypex 6, Woodpex III and Propex Pixi in uniradicular premolars with simulated root resorption.

MATERIALS AND METHODS

For this study, 30 single-rooted permanent premolars were selected. The samples were divided into two groups: 15 teeth with simulated external root resorption and 15 teeth with internal resorption. Each sample was immersed in containers with alginate covering only the root part. Measurements were acquired from the four foramen locators and compared with the working length measurements that have been previously obtained with the direct visual technique and the use of a Digital USB Microscope under 5×. Next, the Shapiro-Wilk normality test was applied. The ANOVA test and -test for related samples were performed, in order to analyze the measurements obtained.

RESULTS

In the teeth with external resorption, no statistically significant differences in the measurements were obtained with Raypex 6 and Propex Pixi foramen locators, but there were statistically significant differences with Root ZX mini and Woodpex III Root Zx mini [Root Zx mini (=0.040) and Woodpex III (=0.000)]. On the other hand, in the samples with internal root resorption, there were no statistically significant differences in measurements with the Root ZX mini, Propex Pixi and Raypex 6 foramen locators. However, there were significant differences with Woodpex III.

CONCLUSIONS

Based on this study, Raypex 6 had the highest accuracy in premolars with simulated external resorption and Root ZX mini was the most accurate in teeth with simulated internal resorption. Furthermore, external root resorption affects the accuracy of foramen locators more than internal resorption.

摘要

引言

本研究的目的是比较四种根管长度测定仪(Root ZX mini、Raypex 6、Woodpex III和Propex Pixi)在模拟根吸收的单根前磨牙中测定工作长度的准确性。

材料与方法

本研究选取了30颗单根恒牙前磨牙。样本分为两组:15颗模拟牙根外吸收的牙齿和15颗模拟牙根内吸收的牙齿。每个样本浸入仅覆盖牙根部分的藻酸盐容器中。从四种根管长度测定仪获取测量值,并与之前通过直接视觉技术和使用5倍数码USB显微镜获得的工作长度测量值进行比较。接下来,应用夏皮罗-威尔克正态性检验。进行方差分析和相关样本的t检验,以分析获得的测量值。

结果

在牙根外吸收的牙齿中,Raypex 6和Propex Pixi根管长度测定仪的测量值无统计学显著差异,但与Root ZX mini和Woodpex III有统计学显著差异[Root Zx mini(P = 0.040)和Woodpex III(P = 0.000)]。另一方面,在牙根内吸收的样本中,Root ZX mini、Propex Pixi和Raypex 6根管长度测定仪的测量值无统计学显著差异。然而,与Woodpex III有显著差异。

结论

基于本研究,Raypex 6在模拟牙根外吸收的前磨牙中准确性最高,Root ZX mini在模拟牙根内吸收的牙齿中最准确。此外,牙根外吸收比牙根内吸收对根管长度测定仪准确性的影响更大。