Mahmoud Okba, Awad Abdelmagied Mawada Hassan, Dandashi Ahmad Hisham, Jasim Bakr Nssaief, Tawfik Kayali Hussam Alddin, Al Shehadat Saaid
Clinical Sciences Department, Faculty of Dentistry, Ajman University, Ajman, UAE.
Department of Preventive and Restorative Dentistry, College of Dental Medicine, University of Sharjah, Sharjah, UAE.
Int J Dent. 2021 May 4;2021:5563426. doi: 10.1155/2021/5563426. eCollection 2021.
This study aimed to validate the accuracy of working length (WL) measurements obtained with the newly introduced Propex IQ apex locator and to compare it with the latest generations of other electronic apex locators, CBCT, and conventional periapical radiographs by using the actual WL measurements obtained by using an endodontics microscope as a reference.
Thirty-five extracted single-rooted human mandibular first premolars with curvatures from 10° to 20° were selected according to the inclusion and exclusion criteria and cut at the cementoenamel junction to achieve a standard reference point for WL determination. The actual WL was obtained by inserting a size-15 k-file in the root canal until the tip of the file was visible under an endodontic microscope. The definitive WL was measured using Propex IQ (Dentsply Sirona), Raypex 6 (VDW Dental), Root ZX (Morita), and Apex ID (Kerr Dental). In addition, radiographic WL was obtained using periapical radiographs and CBCT. One-way ANOVA was used for comparisons of the WL values, with a value < 0.05. The percentage of success of each method for determination of the definitive WL was assessed using cross-tabulation and chi-square tests.
CBCT radiographs and Propex IQ apex locator yielded the most accurate WL measurements in comparison with the actual WL measurements ( < 0.05). Raypex 6, Root ZX, and Apex ID yielded more accurate WL measurements than conventional periapical radiographs ( < 0.05). Periapical radiographs yielded the least accurate WL measurements in comparison with the actual WL values ( < 0.05).
Within the limitations of this study, the Propex IQ apex locator showed higher accuracy than Raypex 6, Root ZX, and Apex ID for WL determination in the root canal. Nevertheless, CBCT radiographs yielded the maximum accuracy for WL measurements.
本研究旨在验证新推出的Propex IQ根尖定位仪获得的工作长度(WL)测量的准确性,并将其与其他新一代电子根尖定位仪、锥形束计算机断层扫描(CBCT)以及传统根尖片进行比较,以使用牙髓显微镜获得的实际WL测量值作为参考。
根据纳入和排除标准,选择35颗拔除的单根人下颌第一前磨牙,其弯曲度为10°至20°,并在牙骨质釉质界处切割,以获得用于确定WL的标准参考点。通过将15号K锉插入根管,直到在牙髓显微镜下可见锉尖,从而获得实际WL。使用Propex IQ(登士柏西诺德)、Raypex 6(VDW牙科)、Root ZX(森田)和Apex ID(克尔牙科)测量最终WL。此外,使用根尖片和CBCT获得影像学WL。采用单因素方差分析比较WL值,P值<0.05。使用交叉表和卡方检验评估每种方法确定最终WL的成功率。
与实际WL测量值相比,CBCT片和Propex IQ根尖定位仪获得的WL测量最准确(P<0.05)。Raypex 6、Root ZX和Apex ID获得的WL测量比传统根尖片更准确(P<0.05)。与实际WL值相比,根尖片获得的WL测量最不准确(P<0.05)。
在本研究的局限性范围内,Propex IQ根尖定位仪在根管WL测定中显示出比Raypex 6、Root ZX和Apex ID更高的准确性。然而,CBCT片在WL测量中获得的准确性最高。