Barakat Reem M, Almohareb Rahaf A, Alharbi Arwa O, Alhazmi Asma, Alomar Reem
Dental Clinics Department, King Abdullah bin Abdulaziz University Hospital, Princess Nourah Bint Abdulrahman University, P.O. Box 84428, Riyadh 11671, Saudi Arabia.
Department of Clinical Dental Sciences, College of Dentistry, Princess Nourah Bint Abdulrahman University, P.O. Box 84428, Riyadh 11671, Saudi Arabia.
J Clin Med. 2024 Sep 2;13(17):5199. doi: 10.3390/jcm13175199.
Determining the correct apical limit for root canal treatment is essential for its success. This study evaluates the accuracy of three electronic apex locators (EALs) in locating the apical constriction (AC) in molar canals. Forty extracted human mandibular molars were scanned using micro-CT, and endodontic access cavities were created. Teeth were mounted in alginate, and three EALs-Root ZX-mini, Root ZX-II, and Sirona integrated apex locator-were used to measure the canal working length in dry canals and with EDTA gel. Micro-CT scans were performed with files in place, and the distance from the AC was calculated. Measurements within 0.1-0.5 mm were categorized as 'close'. Those extending beyond towards the major foramen were categorized as 'beyond', otherwise they were classified as 'far'. Data analysis was conducted with a level of significance set at 5%. Most readings for all EALs were in the 'close' category, with significant differences between devices ( < 0.0001). Root ZX-mini and Root ZX-II had 74.4% and 72.5% 'close' readings, respectively, versus 51% for Sirona integrated. Accuracy did not differ significantly between dry and EDTA-treated canals ( = 0.306). All EALs demonstrated excellent operator reliability (ICC 0.996-1.00). All EALs accurately determined AC, unaffected by lubricants. However, Root ZX-mini and Root ZX-II outperformed Sirona integrated. All EALs showed consistent reliability.
确定根管治疗的正确根尖界限对其成功至关重要。本研究评估了三种电子根尖定位仪(EALs)在磨牙根管中定位根尖狭窄(AC)的准确性。使用微型计算机断层扫描(micro-CT)对40颗拔除的人类下颌磨牙进行扫描,并制备根管入口洞形。将牙齿固定在藻酸盐中,使用三种EALs——Root ZX-mini、Root ZX-II和西诺德综合根尖定位仪——测量干燥根管和使用乙二胺四乙酸(EDTA)凝胶时的根管工作长度。在根管内放置锉后进行微型计算机断层扫描,并计算与根尖狭窄的距离。0.1至0.5毫米范围内的测量结果归类为“接近”。超出根尖孔延伸至主要根尖孔的测量结果归类为“超出”,否则归类为“远”。数据分析的显著性水平设定为5%。所有EALs的大多数读数都属于“接近”类别,不同仪器之间存在显著差异(<0.0001)。Root ZX-mini和Root ZX-II的“接近”读数分别为74.4%和72.5%,而西诺德综合根尖定位仪为51%。干燥根管和EDTA处理根管之间的准确性无显著差异(=0.306)。所有EALs均显示出极佳的操作者可靠性(组内相关系数ICC为0.996至1.00)。所有EALs均能准确确定根尖狭窄,不受润滑剂影响。然而,Root ZX-mini和Root ZX-II的表现优于西诺德综合根尖定位仪。所有EALs均显示出一致的可靠性。