Aljawad Hussein, Lim Hoi-Jeong, Lee Kyungmin Clara
Department of Orthodontics, School of Dentistry, Chonnam National University.
Graduate School of Data Science, Chonnam National University.
J Craniofac Surg. 2023;34(5):1456-1458. doi: 10.1097/SCS.0000000000009076. Epub 2022 Oct 13.
The aim of this study was to evaluate the accuracy of 3-dimensional (3D) facial scans taken with a low-cost facial scanner compared with cone-beam computed tomography (CBCT).
Twenty-five patients had their CBCT scan and 3D facial scan taken on the same day. Facial scans were taken with a low-cost facial scanner (Bellus3D, ARC-1 scanner) in an upright position. The facial scans were imported into Invivo5 software and were integrated with the corresponding CBCT scans. Then, 16 soft tissue landmarks were identified on the CBCT scans and the 3D-integrated facial scans. The 6 vertical, 7 horizontal, 10 oblique linear, and 11 angular anthropometric measurements were obtained and compared between the CBCT scans and 3D-integrated facial scans using paired t test and Bland-Altman plots.
The differences between CBCT scans and 3D-integrated facial scans showed no statistical significance except for 2 vertical measurements. Bland-Altman plots showed that all anthropometric measurements were within the limit of agreement. The differences between CBCT scans and 3D-integrated facial scans were <1.5 mm or 1.5 degrees except for 2 vertical measurements.
Three-dimensional facial scans taken with facial scanner showed a clinical acceptance when compared with CBCT scans.
本研究旨在评估使用低成本面部扫描仪进行的三维(3D)面部扫描与锥形束计算机断层扫描(CBCT)相比的准确性。
25例患者在同一天进行了CBCT扫描和3D面部扫描。使用低成本面部扫描仪(Bellus3D,ARC-1扫描仪)在直立位进行面部扫描。将面部扫描图像导入Invivo5软件,并与相应的CBCT扫描图像进行整合。然后,在CBCT扫描图像和3D整合面部扫描图像上识别16个软组织标志点。获得6项垂直、7项水平、10项斜向线性和11项角度人体测量数据,并使用配对t检验和Bland-Altman图对CBCT扫描图像和3D整合面部扫描图像的数据进行比较。
除2项垂直测量外,CBCT扫描图像和3D整合面部扫描图像之间的差异无统计学意义。Bland-Altman图显示,所有人体测量数据均在一致性界限内。除2项垂直测量外,CBCT扫描图像和3D整合面部扫描图像之间的差异<1.5毫米或1.5度。
与CBCT扫描相比,使用面部扫描仪进行的三维面部扫描具有临床可接受性。