In vivo calibration of a transcutaneous oxygen electrode in adult patients.

作者信息

Gray B J, Heaton R W, Henderson A, Hutchison D C

机构信息

Department of Thoracic Medicine, King's College School of Medicine and Dentistry, London, U.K.

出版信息

Adv Exp Med Biol. 1987;220:75-7. doi: 10.1007/978-1-4613-1927-6_13.

Abstract

Transcutaneous oxygen tension (tcPO2) has been compared with arterial oxygen tension (PaO2 in 14 haemodynamically stable patients in an intensive care unit. Two calibration methods have been compared: (1) "In vitro" calibration, a two point calibration procedure carried out before attachment to the skin. (2) "In vivo" calibration, calibration using a single arterial sample, to recalibrate the upper point after attachment of the electrode to the skin and stabilisation of the electrical output. After "in vitro" calibration the regression equation was given by tcPO2 (mmHg) = 0.58 PaO2 + 13.4 (95% confidence limits +/- 19.6). After "in vivo" calibration, the regression equation for 55 comparisons over the range 50 to 120 mmHg was given by: tcPO2 (mmHg) = 0.98 PaO2 + 1.6 (95% confidence limits +/- 6.6). The "in vivo" calibration method therefore allows a close estimate of PaO2 to be made from tcPO2 values in adult patients providing strict operating criteria observed.