使用经皮电极估计成年受试者的动脉血氧分压。

Estimation of arterial oxygen tension in adult subjects using a transcutaneous electrode.

作者信息

Hutchison D C, Rocca G, Honeybourne D

出版信息

Thorax. 1981 Jun;36(6):473-7. doi: 10.1136/thx.36.6.473.

Abstract

Transcutaneous oxygen tension (tcPO2) measured with a heated electrode was compared with arterial oxygen tension (PaO2) in three groups of adult patients-group A: 20 patients with chronic respiratory disease; group B: eight hypothermic patents studied immediately after cardiopulmonary bypass surgery; and group C: 14 patients in an intensive care unit who had in all 35 comparisons of tcPO2 and PaO2. For group A, the relationship of the two methods is given by the regression equation (in mmHg): tcPO2 = 0.88 PaO2 + 5.0 (standard error = 5.2: r = 0.93: p less than 0.001) and for group C tcPO2 = 0.69 PaO2 + 14.9 (standard error =12.0, r = 0.90: p less than 0.001). A method of in vivo calibration using a single arterial blood sample improved the reliability of the method in group C. In the group B patients there was no correlation between tcPO2 and PaO2 even when supplementary heating was applied to the skin. Continuous monitoring can reveal large fluctuations in arterial PO2 which would be missed by the use of intermittent arterial samples. The transcutaneous electrode can be employed usefully in intensive care monitoring of adult patients and in physiological studies where repeated arterial samples would otherwise be required.

摘要

使用加热电极测量的经皮氧分压(tcPO2)与三组成年患者的动脉氧分压(PaO2)进行了比较:A组:20例慢性呼吸系统疾病患者;B组:8例在体外循环手术后立即进行研究的体温过低患者;C组:14例重症监护病房患者,共进行了35次tcPO2与PaO2的比较。对于A组,两种方法的关系由回归方程(以毫米汞柱为单位)给出:tcPO2 = 0.88PaO2 + 5.0(标准误差 = 5.2:r = 0.93:p < 0.001),对于C组,tcPO2 = 0.69PaO2 + 14.9(标准误差 = 12.0,r = 0.90:p < 0.001)。使用单个动脉血样进行体内校准的方法提高了C组中该方法的可靠性。在B组患者中,即使对皮肤进行补充加热,tcPO2与PaO2之间也没有相关性。连续监测可以揭示动脉PO2的大幅波动,而使用间歇性动脉血样则会遗漏这些波动。经皮电极可有效地用于成年患者的重症监护监测以及需要反复采集动脉血样的生理学研究中。

相似文献

3
In vivo calibration of a transcutaneous oxygen electrode in adult patients.
Adv Exp Med Biol. 1987;220:75-7. doi: 10.1007/978-1-4613-1927-6_13.
4
Transcutaneous oxygen tension during exercise in patients with pulmonary emphysema.
Adv Exp Med Biol. 1987;220:67-70. doi: 10.1007/978-1-4613-1927-6_11.
8
[Transcutaneous measurement of pO2.(author's transl)].
Rev Fr Mal Respir. 1981;9(4):327-35.
10
Reliability of transcutaneous oxygen monitoring of critically ill children in a general pediatric unit.
Clin Pediatr (Phila). 1983 Jun;22(6):431-5. doi: 10.1177/000992288302200607.