Pump-induced hemolysis in a rabbit model of neonatal ECMO.

作者信息

Kress D C, Cohen D J, Swanson D K, Hegge J O, Young J W, Watson K M, Rasmussen P W, Berkoff H A

机构信息

Division of Cardiothoracic Surgery, University of Wisconsin Hospitals and Clinics, Madison.

出版信息

ASAIO Trans. 1987 Jul-Sep;33(3):446-52.

PMID:3675974
Abstract

There is no significant difference in hemolysis between pediatric centrifugal pumps and roller pumps during ECMO of 8 hours duration in rabbits or in comparable 24 hour isolated pump circuits. There is also no significant difference in heat production by either pump. Reports of severe hemolysis using the centrifugal pump in neonatal ECMO are likely due to other factors. Proper cannula selection and positioning are important measures to reduce hemolysis by minimizing circuit pressure and maximizing venous return. Avoidance of inadvertently high rotor speed in the centrifugal pump is also recommended.

摘要

在兔子进行8小时体外膜肺氧合(ECMO)期间或在类似的24小时独立泵循环中,儿科离心泵和滚压泵之间的溶血情况无显著差异。两种泵的产热情况也无显著差异。新生儿ECMO中使用离心泵出现严重溶血的报道可能是由其他因素导致的。正确选择插管并确定其位置是重要措施,可通过最小化回路压力和最大化静脉回流量来减少溶血。还建议避免离心泵转子速度意外过高。