Pisa Justyn F D, Almutairi Walleed H, Mackie Kristy, Stangherlin Daniela, Sulkers Jacob, Hochman Jordan B
University of Manitoba.
Central Speech and Hearing Clinic, Winnipeg, Manitoba, Canada.
Otol Neurotol. 2023 Apr 1;44(4):e211-e215. doi: 10.1097/MAO.0000000000003819. Epub 2023 Feb 10.
In February 2020, Advanced Bionics initiated a field action notice to remove the MS Ultra (V1) implant from circulation. In this study, we quantify a single site's incidence with device failure and examine the relationship between impedance change and declining speech perception.
Retrospective/cohort study.
Tertiary healthcare center.
Forty-nine adult patients (52 devices) were implanted between October 2017 and December 2019, with the following exclusion criteria: <18 years of age, medical/surgical failures, and lack of English-language proficiency.
Diagnostic.
Postoperative speech perception (AzBio sentence test) scores at 12 months postactivation were compared with repeat testing at 6- to 12-month intervals. Degree of change in impedances from 1 month postactivation was analyzed at similar intervals. Device failures were suspected when impedance levels on three or more electrodes dropped to <3.5 kΩ and AzBio scores declined >15%. Device failures were confirmed through analyses completed by the manufacturer.
To date, 21% (11/52) of devices have met all three criteria for a confirmed failure with an average decline in speech perception of 27.7 ± 22%. The average length of time for detection of device failure was 21.9 ± 5 months postimplantation. An additional 7 devices (13%) are currently being monitored for failure but have not yet met full criteria. There was no predictive relationship between degree of impedance change and speech perception scores. Electrodes along the entire array (channels 1-16) can be affected.
Impedance values can be used as a reliable indicator of device malfunction, thus facilitating patient counseling and early intervention. No correlation was identified between impedance changes and speech perception scores.
2020年2月,先进仿生公司发布了一份现场行动通知,要求停止MS Ultra(V1)植入物的流通。在本研究中,我们对单个部位的设备故障发生率进行了量化,并研究了阻抗变化与言语感知能力下降之间的关系。
回顾性/队列研究。
三级医疗中心。
49例成年患者(52个设备)于2017年10月至2019年12月期间植入,排除标准如下:年龄<18岁、医疗/手术失败以及英语不熟练。
诊断。
激活后12个月的术后言语感知(AzBio句子测试)分数与每隔6至12个月的重复测试结果进行比较。在相似的时间间隔分析激活后1个月起的阻抗变化程度。当三个或更多电极的阻抗水平降至<3.5 kΩ且AzBio分数下降>15%时,怀疑设备出现故障。通过制造商完成的分析来确认设备故障。
迄今为止,21%(11/52)的设备已满足确认故障的所有三项标准,言语感知平均下降27.7±22%。检测到设备故障的平均时间为植入后21.9±5个月。另外7个设备(13%)目前正在监测是否出现故障,但尚未达到全部标准。阻抗变化程度与言语感知分数之间没有预测关系。整个阵列(通道1 - 16)上的电极都可能受到影响。
阻抗值可作为设备故障的可靠指标,从而便于患者咨询和早期干预。未发现阻抗变化与言语感知分数之间存在相关性。