Prunaretty Jessica, Mir Nicolas, Tilhac Anaïs, Gouillou Maureen, Azria David, Fenoglietto Pascal, Bourgier Céline
Department of Radiation Oncology, Montpellier Cancer Institute, 34298 Montpellier, France.
J Clin Med. 2023 Jan 21;12(3):862. doi: 10.3390/jcm12030862.
In our hospital, a TrueBeam linear accelerator and the PerfectPitch 6-degrees-of-freedom (6-DOF) couch (Varian), with 7 mm margins, are used for volumetric modulated arc therapy (VMAT) of breast cancer (BC). This study tested whether a 3-degrees-of-freedom (3-DOF) couch, i.e., without rotation compensation (such as the Halcyon couch), affected dose metrics. A total of 446 daily extended cone beam computed tomography (CBCT) data of 20 patients who received VMAT for BC were used to recalculate the treatment plans with the session registration (6-DOF) and a simulated matching with 3-DOF. The initial plan provided significantly better coverage for internal mammary chain and clavicular lymph node clinical target volumes (CTVs) than the 6-DOF and 3-DOF CBCT plans. The volumes receiving 110% of the prescribed dose (V110%) were increased for all CTVs with the 6-DOF and 3-DOF CBCT plans, but the difference was significant only for the breast/chest wall CTV ( < 0.05; paired samples -test). Protection of the heart and lungs was comparable among plans. The dose volume histograms based on the 6-DOF and 3-DOF data were similar for CTVs and organs at risk. Therefore, with a 7 mm margin, VMAT and a 3-DOF couch can be used for BC treatment without any compromise in delivery accuracy.
在我们医院,使用TrueBeam直线加速器和PerfectPitch六自由度(6-DOF)治疗床(瓦里安公司),设置7毫米的边界,用于乳腺癌(BC)的容积调强弧形放疗(VMAT)。本研究测试了三自由度(3-DOF)治疗床,即没有旋转补偿功能的治疗床(如Halcyon治疗床)是否会影响剂量指标。总共使用了20例接受BC的VMAT治疗患者的446份每日扩展锥束计算机断层扫描(CBCT)数据,分别使用疗程配准(6-DOF)和模拟的3-DOF匹配来重新计算治疗计划。初始计划对乳腺内链和锁骨上淋巴结临床靶区(CTV)的覆盖明显优于6-DOF和3-DOF的CBCT计划。对于所有CTV,6-DOF和3-DOF的CBCT计划中接受规定剂量110%(V110%)的体积均增加,但仅乳腺/胸壁CTV的差异具有统计学意义(<0.05;配对样本t检验)。各计划对心脏和肺部的保护相当。基于6-DOF和3-DOF数据的剂量体积直方图在CTV和危及器官方面相似。因此,设置7毫米的边界时,VMAT和3-DOF治疗床可用于BC治疗,而不会在治疗准确性上有任何妥协。