Preclinical Assessment of Tissue Effects by Gastrointestinal Endoscope Tip Temperature.

作者信息

Jäger Luise, Morales-Orcajo Enrique, Gager Anna, Bader Anke, Dillinger Anja, Blutke Andreas

机构信息

Ambu Innovation GmbH, Augsburg, Germany.

Institute of Veterinary Pathology, Center for Clinical Veterinary Medicine, Ludwig-Maximilians Universitaet München, Munich, Germany.

出版信息

Curr Ther Res Clin Exp. 2023 Jan 26;98:100693. doi: 10.1016/j.curtheres.2023.100693. eCollection 2023.

Abstract

BACKGROUND

Endoscope tips are heated by their electrical and illuminating components. During the procedure, they might get in close or even direct contact with intestinal tissues.

OBJECTIVE

To assess endoscope tip and tissue temperature as well as histopathologic changes of gastrointestinal (GI) tissues when exposed to the heated tip of GI endoscopes.

METHODS

The endoscope tip temperatures of four GI endoscopes were measured for 30 minutes in a temperature-controlled chamber. The temperature of ex vivo porcine GI tissues was measured for 5-, 15-, and 120-minute exposure to endoscope tips within a climate chamber to control environmental factors (simulation of fever as worst-case). Exposed tissues were histopathologically examined afterward. Control samples included untreated mucosa, tissue samples exposed to endoscope tips for 120 minutes, as well as tissue samples thermally coagulated with a bipolar high-frequency probe.

RESULTS

Actual endoscope tip temperatures of 59 to 86°C, dependent on the endoscope type, were measured. After 10 to 15 minutes, the maximum temperatures were reached. Maximum tissue temperatures of 44 to 46°C for 5 and 15 minutes, as well as up to 50°C for 120 minutes, were recorded dependent on tissue and endoscope type. No direct heat-induced histopathologic tissue alterations were observed in the 5- and 15-minute samples.

CONCLUSIONS

Both clinically relevant and a worst-case control were tested. Even though elevated temperatures were recorded, no heat-related tissue alterations were detected. This overall supports the safety profile of GI endoscopy; however, the study findings are limited by the ex vivo setting (no metabolic tissue alterations accessible, no blood flow) and small sample number.

摘要

背景

内镜尖端会因其电气和照明部件而发热。在操作过程中,它们可能会与肠道组织近距离甚至直接接触。

目的

评估胃肠道(GI)内镜加热尖端接触胃肠道组织时的内镜尖端温度和组织温度,以及胃肠道组织的组织病理学变化。

方法

在温度控制箱中对4种胃肠道内镜的尖端温度进行30分钟测量。在气候箱中,将离体猪胃肠道组织暴露于内镜尖端5分钟、15分钟和120分钟,测量其温度,以控制环境因素(模拟最坏情况下的发热)。之后对暴露的组织进行组织病理学检查。对照样本包括未处理的黏膜、暴露于内镜尖端120分钟的组织样本,以及用双极高频探头热凝的组织样本。

结果

测量得到的内镜尖端实际温度为59至86°C,取决于内镜类型。10至15分钟后达到最高温度。根据组织和内镜类型,记录到5分钟和15分钟时组织最高温度为44至46°C,120分钟时最高可达50°C。在5分钟和15分钟的样本中未观察到直接的热诱导组织病理学改变。

结论

对临床相关情况和最坏情况对照均进行了测试。尽管记录到温度升高,但未检测到与热相关的组织改变。这总体上支持了胃肠道内镜检查的安全性;然而,该研究结果受离体设置(无法检测代谢性组织改变,无血流)和样本量小的限制。