Is the effectiveness of cardiac ventricular defibrillation dependent upon polarity?

作者信息

Schuder J C, Stoeckle H, McDaniel W C, Dbeis M

机构信息

Department of Surgery, University of Missouri, Columbia 65212.

出版信息

Med Instrum. 1987 Oct;21(5):262-5.

PMID:3683252
Abstract

Studies involving 240 fibrillation-defibrillation episodes via epicardial patch electrodes in 21-27-kg dogs and 480 transthoracic episodes in 100-kg calves are reported. In dogs, 120 episodes involving shocks by a 3.7-A, 5-ms unidirectional rectangular wave of one polarity were interlaced with 120 similar episodes of the reverse polarity. When the upper right ventricular patch was positive with respect to the left ventricular apex patch, 85% of the episodes yielded defibrillation; 78% defibrillated with the reverse polarity. In one series in calves, 120 episodes involving shocks by a 42-A, 4-ms unidirectional rectangular wave of one polarity were interlaced with 120 similar episodes involving the reverse polarity. In a second series, 59-A, 4-ms shocks were employed. In the 42-A series, 38% of the episodes were successful when the upper right electrode was positive with respect to the electrode over the apex, and 28% were successful with the reverse polarity. Corresponding results in the 59-A series were 80% and 68%, respectively. On an unpaired basis, the three p values were 0.19, 0.14, and 0.04, respectively. On a paired basis, all three comparisons yielded significant differences (p less than 0.05). We conclude that in these cases, at least, there is a moderate dependence upon electrode polarity and that our results appear to warrant clinical studies.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验