Evaluation of patient organ doses from kilovoltage cone-beam CT imaging in radiation therapy.

作者信息

Özseven Alper, Dirican Bahar

机构信息

Suleyman Demirel University, Medical Faculty, Isparta, Turkey.

University of Health Sciences, Gulhane Medical Faculty, Ankara, Turkey.

出版信息

Rep Pract Oncol Radiother. 2021 Apr 14;26(2):251-258. doi: 10.5603/RPOR.a2021.0038. eCollection 2021.

Abstract

BACKGROUND

Currently, CBCT system is an indispensable component of radiation therapy units. Because of that, it is important in treatment planning and diagnosis. CBCT is also an crucial tool for patient positioning and verification in image-guided radiation therapy (IGRT). Therefore, it is critical to investigate the patient organ doses arising from CBCT imaging. The purpose of this study is to evaluate patient organ doses and effective dose to patients from three different protocols of Elekta Synergy XVI system for kV CBCT imaging examinations in image guided radiation therapy.

MATERIALS AND METHODS

Organ dose measurements were done with thermoluminescent dosimeters in Alderson RA NDO male phantom for head & neck (H&N), chest and pelvis protocols of the Elekta Synergy XVI kV CBCT system. From the measured organ dose, effective dose to patients were calculated according to the International Commission on Radiological Protection 103 report recommendations.

RESULTS

For H&N, chest and pelvis scans, the organ doses were in the range of 0.03-3.43 mGy, 6.04-22.94 mGy and 2.5-25.28 mGy, respectively. The calculated effective doses were 0.25 mSv, 5.56 mSv and 4.72 mSv, respectively.

CONCLUSION

The obtained results were consistent with the most published studies in the literature. Although the doses to patient organs from the kV CBCT system were relatively low when compared with the prescribed treatment dose, the amount of delivered dose should be monitored and recorded carefully in order to avoid secondary cancer risk, especially in pediatric examinations.

摘要

背景

目前,CBCT系统是放射治疗设备不可或缺的组成部分。正因如此,它在治疗计划和诊断中至关重要。CBCT也是图像引导放射治疗(IGRT)中患者定位和验证的关键工具。因此,研究CBCT成像产生的患者器官剂量至关重要。本研究的目的是评估在图像引导放射治疗中,Elekta Synergy XVI系统用于kV CBCT成像检查的三种不同方案对患者的器官剂量和有效剂量。

材料与方法

在Alderson RA NDO男性体模中,使用热释光剂量计对Elekta Synergy XVI kV CBCT系统的头颈部(H&N)、胸部和骨盆方案进行器官剂量测量。根据国际放射防护委员会103号报告的建议,从测量的器官剂量计算患者的有效剂量。

结果

对于头颈部、胸部和骨盆扫描,器官剂量分别在0.03 - 3.43 mGy、6.04 - 22.94 mGy和2.5 - 25.28 mGy范围内。计算得到的有效剂量分别为0.25 mSv、5.56 mSv和4.72 mSv。

结论

获得的结果与文献中大多数已发表的研究一致。尽管与规定的治疗剂量相比,kV CBCT系统对患者器官的剂量相对较低,但为避免二次癌症风险,尤其是在儿科检查中,应仔细监测和记录所输送的剂量。