双能CT的虚拟单色图像作为单能CT的替代方法:使用针对不同采集技术的可检测性指数进行性能比较。

Virtual monochromatic images of dual-energy CT as an alternative to single-energy CT: performance comparison using a detectability index for different acquisition techniques.

作者信息

Kawashima Hiroki, Ichikawa Katsuhiro, Ueta Hiroshi, Takata Tadanori, Mitsui Wataru, Nagata Hiroji

机构信息

Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, 920-0942, Japan.

Radiology Division, Kanazawa University Hospital, 13-1 Takara-Machi, Kanazawa, 920-8641, Japan.

出版信息

Eur Radiol. 2023 Aug;33(8):5752-5760. doi: 10.1007/s00330-023-09491-6. Epub 2023 Mar 9.

Abstract

OBJECTIVES

To investigate the performance of virtual monochromatic (VM) images with the same dose and iodine contrast as those for single-energy (SE) images using five dual-energy (DE) scanners with DE techniques: two generations of fast kV switching (FKS), two generations of dual source (DS), and one split filter (SF).

METHODS

A water-bath phantom with a diameter of 300 mm, which contains one rod-shaped phantom made of a material equivalent to soft-tissue and two rod-shaped phantoms made of diluted iodine (2 and 12 mg/mL), was scanned using both SE (120, 100, and 80 kV) and DE techniques with the same CT dose index in each scanner. The VM energy at which the CT number of the iodine rod is closest to that of each SE tube voltage was determined as the equivalent energy (Eeq). A detectability index (d') was calculated from the noise power spectrum, the task transfer functions, and a task function corresponding to each rod. The percentage of the d' value of the VM image to that of the corresponding SE image was calculated for performance comparison.

RESULTS

The average percentages of d' of FKS1, FKS2, DS1, DS2, and SF were 84.6%, 96.2%, 94.3%, 107%, and 104% for 120 kV-Eeq; 75.9%, 91.2%, 88.2%, 99.2%, and 82.6% for 100 kV-Eeq; 71.6%, 88.9%, 82.6%, 85.2%, and 62.3% for 80 kV-Eeq, respectively.

CONCLUSION

The performance of VM images was on the whole inferior to that of SE images especially at low equivalent energy levels, depending on the DE techniques and their generations.

KEY POINTS

• This study evaluated the performance of VM images with the same dose and iodine contrast as those for SE images using five DE scanners. • The performance of VM images varied with the DE techniques and their generations and was mostly inferior at low equivalent energy levels. • The results highlight the importance of distribution of available dose over the two energy levels and spectral separation for the performance improvement of VM images.

摘要

目的

使用五台采用双能(DE)技术的双能扫描仪研究虚拟单色(VM)图像的性能,这些扫描仪的剂量和碘对比剂与单能(SE)图像相同,DE技术包括两代快速千伏切换(FKS)、两代双源(DS)和一种分体式滤光器(SF)。

方法

使用一台直径300毫米的水浴模体进行扫描,该模体包含一个由软组织等效材料制成的棒状模体以及两个由稀释碘(2和12毫克/毫升)制成的棒状模体,在每台扫描仪中使用SE(120、100和80千伏)和DE技术,且CT剂量指数相同。将碘棒的CT值最接近每个SE管电压时的VM能量确定为等效能量(Eeq)。根据噪声功率谱、任务传递函数以及与每个棒对应的任务函数计算可探测性指数(d')。计算VM图像的d'值相对于相应SE图像d'值的百分比以进行性能比较。

结果

对于120 kV-Eeq,FKS1、FKS2、DS1、DS2和SF的d'平均百分比分别为84.6%、96.2%、94.3%、107%和104%;对于100 kV-Eeq,分别为75.9%、91.2%、88.2%、99.2%和82.6%;对于80 kV-Eeq,分别为71.6%、88.9%、82.6%、85.2%和62.3%。

结论

VM图像的性能总体上低于SE图像,尤其是在低等效能量水平时,这取决于DE技术及其代数。

关键点

• 本研究使用五台DE扫描仪评估了与SE图像具有相同剂量和碘对比剂的VM图像的性能。• VM图像的性能随DE技术及其代数而变化,且在低等效能量水平时大多较差。• 结果突出了在两个能量水平上分配可用剂量以及光谱分离对于改善VM图像性能的重要性。