四种用于腹部成像的双能 CT 平台的性能:基于体模数据的基于任务的图像质量评估。

Performance of four dual-energy CT platforms for abdominal imaging: a task-based image quality assessment based on phantom data.

机构信息

Department of Medical Imaging, CHU Nimes, Medical Imaging Group Nimes, EA 2415, Univ Montpellier, Bd Prof Robert Debré, 30029, Nîmes Cedex 9, France.

Department of Radiology, Hospices Civils de Lyon, 69500, Lyon, France.

出版信息

Eur Radiol. 2021 Jul;31(7):5324-5334. doi: 10.1007/s00330-020-07671-2. Epub 2021 Jan 15.

Abstract

OBJECTIVES

To compare the spectral performance of dual-energy CT (DECT) platforms using task-based image quality assessment based on phantom data.

MATERIALS AND METHODS

Two CT phantoms were scanned on four DECT platforms: fast kV-switching CT (KVSCT), split filter CT (SFCT), dual-source CT (DSCT), and dual-layer CT (DLCT). Acquisitions on each phantom were performed using classical parameters of abdomen-pelvic examination and a CTDI at 10 mGy. Noise power spectrum (NPS) and task-based transfer function (TTF) were evaluated from 40 to 140 keV of virtual monoenergetic images. A detectability index (d') was computed to model the detection task of two contrast-enhanced lesions as function of keV.

RESULTS

The noise magnitude decreased from 40 to 70 keV for all DECT platforms, and the highest noise magnitude values were found for KVSCT and SFCT and the lowest for DSCT and DLCT. The average NPS spatial frequency shifted towards lower frequencies as the energy level increased for all DECT platforms, smoothing the image texture. TTF values decreased with the increase of keV deteriorating the spatial resolution. For both simulated lesions, higher detectability (d' value) was obtained at 40 keV for DLCT, DSCT, and SFCT but at 70 keV for KVSCT. The detectability of both simulated lesions was highest for DLCT and DSCT.

CONCLUSION

Highest detectability was found for DLCT for the lowest energy levels. The task-based image quality assessment used for the first time for DECT acquisitions showed the benefit of using low keV for the detection of contrast-enhanced lesions.

KEY POINTS

• Detectability of both simulated contrast-enhanced lesions was higher for dual-layer CT for the lowest energy levels. • The image noise increased and the image texture changed for the lowest energy levels. • The detectability of both simulated contrast-enhanced lesions was highest at 40 keV for all dual-energy CT platforms except for fast kV-switching platform.

摘要

目的

基于体模数据的基于任务的图像质量评估,比较双能 CT(DECT)平台的光谱性能。

材料和方法

对四个 DECT 平台(快速千伏切换 CT(KVSCT)、分体式滤过 CT(SFCT)、双源 CT(DSCT)和双层 CT(DLCT))上的两个 CT 体模进行了扫描。使用腹部-骨盆检查的经典参数和 10 mGy 的 CTDI 对每个体模进行采集。从虚拟单能图像的 40 到 140keV 评估噪声功率谱(NPS)和基于任务的传递函数(TTF)。计算出检测两个对比增强病变的检测指数(d')作为 keV 的函数。

结果

所有 DECT 平台的噪声幅度均从 40 降低到 70keV,KVSCT 和 SFCT 的噪声幅度最高,DSCT 和 DLCT 的噪声幅度最低。对于所有 DECT 平台,平均 NPS 空间频率随着能量水平的增加而向低频移动,从而平滑了图像纹理。随着 keV 的增加,TTF 值下降,空间分辨率恶化。对于两个模拟病变,DLCT、DSCT 和 SFCT 的 40keV 时可获得更高的检测能力(d'值),而 KVSCT 的 70keV 时可获得更高的检测能力。对于两种模拟病变,DLCT 和 DSCT 的检测能力最高。

结论

对于最低能级,DLCT 的检测能力最高。首次用于 DECT 采集的基于任务的图像质量评估显示了使用低 keV 检测对比增强病变的优势。

要点

• 对于最低能级,双层 CT 对两种模拟对比增强病变的检测能力更高。

• 图像噪声增加,图像纹理随最低能级变化。

• 除快速千伏切换平台外,所有双能 CT 平台的两种模拟对比增强病变的检测能力在 40keV 时最高。