Department of Translational Research and the New Technologies in Medicine and Surgery, University of Pisa, 56126 Pisa, Italy.
Hospital Management, University Hospital of Pisa, 56126 Pisa, Italy.
Int J Environ Res Public Health. 2023 Feb 28;20(5):4284. doi: 10.3390/ijerph20054284.
Improving the cleaning and disinfection of high-touch surfaces is one of the core components of reducing healthcare-associated infections. The effectiveness of an enhanced protocol applying UV-C irradiation for terminal room disinfection between two successive patients was evaluated. Twenty high-touch surfaces in different critical areas were sampled according to ISO 14698-1, both immediately pre- and post-cleaning and disinfection standard operating protocol (SOP) and after UV-C disinfection (160 sampling sites in each condition, 480 in total). Dosimeters were applied at the sites to assess the dose emitted. A total of 64.3% (103/160) of the sampling sites tested after SOP were positive, whereas only 17.5% (28/160) were positive after UV-C. According to the national hygienic standards for health-care setting, 9.3% (15/160) resulted in being non-compliant after SOP and only 1.2% (2/160) were non-compliant after UV-C disinfection. Operation theaters was the setting that resulted in being less compliant with the standard limit (≤15 colony-forming unit/24 cm) after SOP (12%, 14/120 sampling sites) and where the UV-C treatment showed the highest effectiveness (1.6%, 2/120). The addition of UV-C disinfection to the standard cleaning and disinfection procedure had effective results in reducing hygiene failures.
提高高接触表面的清洁和消毒水平是降低医源性感染的核心措施之一。本研究评估了一种增强型方案的效果,该方案在连续两位患者之间使用紫外线 C 照射对终末病房进行消毒。根据 ISO 14698-1,对不同关键区域的 20 个高接触表面进行采样,分别在清洁和消毒标准操作程序(SOP)之前和之后以及紫外线 C 消毒后(每种情况下 160 个采样点,总共 480 个)进行采样。在采样点应用剂量计来评估发射剂量。SOP 后,共有 64.3%(103/160)的采样点检测呈阳性,而紫外线 C 消毒后只有 17.5%(28/160)呈阳性。根据国家卫生标准,SOP 后有 9.3%(15/160)不符合标准,紫外线 C 消毒后只有 1.2%(2/160)不符合标准。手术室是 SOP 后不符合标准(≤15 菌落形成单位/24cm)的比例最高的环境(12%,14/120 个采样点),而紫外线 C 处理显示出最高的有效性(1.6%,2/120)。将紫外线 C 消毒添加到标准清洁和消毒程序中,可有效降低卫生不合格率。