Suppr超能文献

不同传输技术中小质子束的侧向剂量分布和输出因子的全面研究。

Comprehensive investigation of lateral dose profile and output factor measurements in small proton fields from different delivery techniques.

机构信息

University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl-von-Ossietzky University Oldenburg, Oldenburg, Germany.

Department of Radiation Oncology, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.

出版信息

Med Phys. 2023 Jul;50(7):4546-4561. doi: 10.1002/mp.16357. Epub 2023 Apr 3.

Abstract

BACKGROUND AND PURPOSE

As a part of the commissioning and quality assurance in proton beam therapy, lateral dose profiles and output factors have to be acquired. Such measurements can be performed with point detectors and are especially challenging in small fields or steep lateral penumbra regions as the detector's volume effect may lead to perturbations. To address this issue, this work aims to quantify and correct for such perturbations of six point detectors in small proton fields created via three different delivery techniques.

METHODS

Lateral dose profile and output measurements of three proton beam delivery techniques (pencil beam scanning, pencil beam scanning combined with collimators, passive scattering with collimators) were performed using high-resolution EBT3 films, a PinPoint 3D 31022 ionization chamber, a microSilicon diode 60023 and a microDiamond detector 60019 (all PTW Freiburg, Germany). Detector specific lateral dose response functions K(x,y) acting as the convolution kernel transforming the undisturbed dose distribution D(x,y) into the measured signal profiles M(x,y) were applied to quantify perturbations of the six investigated detectors in the proton fields and correct the measurements. A signal theoretical analysis in Fourier space of the dose distributions and detector's K(x,y) was performed to aid the understanding of the measurement process with regard to the combination of detector choice and delivery technique.

RESULTS

Quantification of the lateral penumbra broadening and signal reduction at the fields center revealed that measurements in the pencil beam scanning fields are only compromised slightly even by large volume ionization chambers with maximum differences in the lateral penumbra of 0.25 mm and 4% signal reduction at the field center. In contrast, radiation techniques with collimation are not accurately represented by the investigated detectors as indicated by a penumbra broadening up to 1.6 mm for passive scattering with collimators and 2.2 mm for pencil beam scanning with collimators. For a 3 mm diameter collimator field, a signal reduction at field center between 7.6% and 60.7% was asserted. Lateral dose profile measurements have been corrected via deconvolution with the corresponding K(x,y) to obtain the undisturbed D(x,y). Corrected output ratios of the passively scattered collimated fields obtained for the microDiamond, microSilicon and PinPoint 3D show agreement better than 0.9% (one standard deviation) for the smallest field size of 3 mm.

CONCLUSION

Point detector perturbations in small proton fields created with three delivery techniques were quantified and found to be especially pronounced for collimated small proton fields with steep dose gradients. Among all investigated detectors, the microSilicon diode showed the smallest perturbations. The correction strategies based on detector's K(x,y) were found suitable for obtaining unperturbed lateral dose profiles and output factors. Approximation of K(x,y) by considering only the geometrical averaging effect has been shown to provide reasonable prediction of the detector's volume effect. The findings of this work may be used to guide the choice of point detectors in various proton fields and to contribute toward the development of a code of practice for small field proton dosimetry.

摘要

背景与目的

作为质子束治疗中的调试和质量保证的一部分,必须获取横向剂量分布和输出因子。此类测量可以使用点探测器进行,尤其是在小场或陡峭的横向半影区域中,由于探测器的体积效应可能会导致干扰,因此具有挑战性。为了解决这个问题,这项工作旨在定量并纠正三种不同递送技术产生的小质子场中六个点探测器的这种干扰。

方法

使用高分辨率 EBT3 胶片、PinPoint 3D 31022 电离室、microSilicon 二极管 60023 和 microDiamond 探测器 60019(均为德国 PTW Freiburg)对三种质子束递送技术(铅笔束扫描、铅笔束扫描结合准直器、被动散射结合准直器)的横向剂量分布和输出进行测量。探测器特定的横向剂量响应函数 K(x,y) 作为卷积核,将未受干扰的剂量分布 D(x,y) 转换为测量的信号分布 M(x,y),用于定量研究六个研究探测器在质子场中的干扰,并校正测量结果。在傅立叶空间中对剂量分布和探测器的 K(x,y) 进行信号理论分析,以帮助理解探测器选择和递送技术组合的测量过程。

结果

对横向半影展宽和场中心信号降低的定量表明,即使使用最大差异为 0.25 毫米的大体积电离室,铅笔束扫描场中的测量也仅受到轻微影响,场中心的信号降低最大为 4%。相比之下,具有准直的放射技术不能由所研究的探测器准确表示,这表明被动散射结合准直的半影展宽高达 1.6 毫米,铅笔束扫描结合准直的半影展宽高达 2.2 毫米。对于 3 毫米直径的准直器场,断言场中心的信号降低介于 7.6%和 60.7%之间。通过与相应的 K(x,y) 反卷积对横向剂量分布进行了校正,以获得未受干扰的 D(x,y)。对于最小的 3 毫米场尺寸,被动散射准直场的校正后输出比获得了微金刚石、微硅和 PinPoint 3D 的一致性好于 0.9%(一个标准差)。

结论

在三种递送技术下创建的小质子场中,对点探测器的干扰进行了量化,发现对于具有陡峭剂量梯度的准直小质子场,干扰尤其明显。在所研究的所有探测器中,微硅二极管的干扰最小。基于探测器 K(x,y) 的校正策略被发现适合获取未受干扰的横向剂量分布和输出因子。仅考虑几何平均效应来近似 K(x,y),已被证明可以合理地预测探测器的体积效应。这项工作的结果可用于指导各种质子场中选择点探测器,并为制定小场质子剂量学实践准则做出贡献。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验