基于双层探测器的光谱CT定量参数在识别缺血性卒中中的诊断价值。

The diagnostic value of quantitative parameters on dual-layer detector-based spectral CT in identifying ischaemic stroke.

作者信息

Huang Jian, Chen Jinghua, Wang Ximing, Hao Ling, Zhang Jinfeng, Zhang Xiaohui, Sheng Zhihong, Liu Kefu

机构信息

Department of Radiology, Taicang Affiliated Hospital of Nanjing University of Traditional Chinese Medicine, Taicang, Jiangsu, China.

Department of Radiology, Taicang Hospital, The Affiliated Hospital of Jiangsu Vocational College of Medicine, Yancheng, Jiangsu, China.

出版信息

Front Neurol. 2023 Feb 23;14:1056941. doi: 10.3389/fneur.2023.1056941. eCollection 2023.

Abstract

OBJECTIVE

To investigate the diagnostic value of quantitative parameters of spectral computed tomography (CT) in ischaemic stroke areas.

METHODS

The medical records of 57 patients with acute ischaemic stroke (AIS) who underwent plain computed tomography (CT) head scans, CT angiography (CTA), and CT perfusion (CTP) were retrospectively reviewed. The ischaemic areas (including the core infarct area and penumbra) and non-ischaemic areas in each patient were quantitatively analyzed using F-STROKE software. Two independent readers measured the corresponding values of the spectroscopic quantitative parameters (effective atomic number [Zeff value], iodine density value, and iodine-no-water value) in the ischaemic area and contralateral normal area alone. The differences in spectroscopic quantitative parameters between the two groups were compared, and their diagnostic efficacy was obtained.

RESULTS

The Zeff, iodine-no-water value, and iodine density value of the ischaemic area all showed significant lower than those of non-ischaemic tissue ( < 0.001). For differentiating the ischaemic area from non-ischaemic tissue, the area under the curve (AUC) of the Zeff value reached 0.869 (cut-off value: 7.385; sensitivity: 93.0%; specificity: 70.2%), the AUC of the iodine density value reached 0.932 (cut-off value: 0.235; sensitivity: 91.2%; specificity: 82.5%), and the AUC of the iodine-no-water value reached 0.922 (cut-off value: 0.205; sensitivity: 96.5%; specificity: 78.9%).

CONCLUSION

The study showed the spectral CT would be a potential novel rapid method for identifying AIS. The spectral CT quantitative parameters (Zeff, iodine density values, and iodine-no-water values) can effectively differentiate the ischaemic area from non-ischaemic tissue in stroke patients.

摘要

目的

探讨光谱计算机断层扫描(CT)定量参数在缺血性卒中区域的诊断价值。

方法

回顾性分析57例接受头颅平扫CT、CT血管造影(CTA)及CT灌注(CTP)检查的急性缺血性卒中(AIS)患者的病历资料。使用F-STROKE软件对每位患者的缺血区域(包括梗死核心区和半暗带)及非缺血区域进行定量分析。两名独立阅片者仅测量缺血区域及对侧正常区域的光谱定量参数(有效原子序数[Zeff值]、碘密度值和碘水比)的对应值。比较两组光谱定量参数的差异,并得出其诊断效能。

结果

缺血区域的Zeff值、碘水比和碘密度值均显著低于非缺血组织(<0.001)。对于区分缺血区域与非缺血组织,Zeff值的曲线下面积(AUC)达到0.869(截断值:7.385;敏感度:93.0%;特异度:70.2%),碘密度值的AUC达到0.932(截断值:0.235;敏感度:91.2%;特异度:82.5%),碘水比的AUC达到0.922(截断值:0.205;敏感度:96.5%;特异度:78.9%)。

结论

该研究表明光谱CT可能是一种潜在的新型快速识别急性缺血性卒中的方法。光谱CT定量参数(Zeff值、碘密度值和碘水比)可有效区分卒中患者的缺血区域与非缺血组织。