Diagnostic accuracy for CZT gamma camera compared to conventional gamma camera technique with myocardial perfusion single-photon emission computed tomography: Assessment of myocardial infarction and function.

机构信息

Department of Clinical Physiology, Skåne University Hospital, Lund University, Lund, Sweden.

Radiation Physics, Department of Hematology, Oncology and Radiation Physics, Skåne University Hospital, Lund, Sweden.

出版信息

J Nucl Cardiol. 2023 Oct;30(5):1935-1946. doi: 10.1007/s12350-022-03185-0. Epub 2023 Mar 13.

Abstract

BACKGROUND

The solid-state cadmium-zinc-telluride (CZT) gamma camera for myocardial perfusion single-photon emission computed tomography (MPS) has theoretical advantages compared to the conventional gamma camera technique. This includes more sensitive detectors and better energy resolution. We aimed to explore the diagnostic performance of gated MPS with a CZT gamma camera compared to a conventional gamma camera for detection of myocardial infarct (MI) and assessment of left ventricular (LV) volumes and ejection fraction (LVEF), using cardiac magnetic resonance (CMR) as the reference method.

METHODS

Seventy-three patients (26% female) with known or suspected chronic coronary syndrome were examined with gated MPS using both a CZT gamma camera and a conventional gamma camera as well as with CMR. Presence and extent of MI on MPS and late gadolinium enhancement (LGE) CMR was evaluated. For LV volumes, LVEF and LV mass, gated MPS images and cine CMR images were evaluated.

RESULTS

MI was found in 42 patients on CMR. The overall sensitivity, specificity, positive and negative predictive values for the CZT and the conventional gamma camera were the same (67%, 100%, 100% and 69%). For infarct size > 3% on CMR, the sensitivity was 82% for the CZT and 73% for the conventional gamma camera, respectively. LV volumes were significantly underestimated by MPS compared to CMR (P ≤ .002 for all measures). The underestimation was slightly less pronounced for the CZT compared to the conventional gamma camera (2-10 mL, P ≤ .03 for all measures). For LVEF, however, accuracy was high for both gamma cameras.

CONCLUSION

Differences between a CZT and a conventional gamma camera for detection of MI and assessment of LV volumes and LVEF are small and do not appear to be clinically significant.

摘要

背景

与传统的伽马相机技术相比,用于心肌灌注单光子发射计算机断层扫描(MPS)的固态碲化镉锌(CZT)伽马相机具有理论优势。这包括更敏感的探测器和更好的能量分辨率。我们旨在探索门控 MPS 使用 CZT 伽马相机与传统伽马相机相比在检测心肌梗死(MI)和评估左心室(LV)容积和射血分数(LVEF)方面的诊断性能,以心脏磁共振(CMR)作为参考方法。

方法

73 例(26%为女性)已知或疑似慢性冠状动脉综合征的患者使用 CZT 伽马相机和传统伽马相机进行门控 MPS 检查,并进行 CMR 检查。评估 MPS 上 MI 的存在和范围以及延迟钆增强(LGE)CMR。对于 LV 容积、LVEF 和 LV 质量,评估门控 MPS 图像和电影 CMR 图像。

结果

CMR 发现 42 例患者存在 MI。CZT 和传统伽马相机的总体敏感性、特异性、阳性和阴性预测值相同(67%、100%、100%和 69%)。对于 CMR 上梗死面积>3%,CZT 的敏感性为 82%,传统伽马相机的敏感性为 73%。与 CMR 相比,MPS 明显低估了 LV 容积(所有测量值 P ≤.002)。与传统伽马相机相比,CZT 的低估程度略低(2-10 毫升,所有测量值 P ≤.03)。然而,对于 LVEF,两种伽马相机的准确性都很高。

结论

CZT 和传统伽马相机在检测 MI 以及评估 LV 容积和 LVEF 方面的差异很小,且似乎无临床意义。