利用脉络膜上视网膜假体使用者的眼球运动估计光幻视位置。

Estimating Phosphene Locations Using Eye Movements of Suprachoroidal Retinal Prosthesis Users.

机构信息

Bionics Institute, East Melbourne, VIC, Australia.

Medical Bionics Department, University of Melbourne, Melbourne, VIC, Australia.

出版信息

Transl Vis Sci Technol. 2023 Mar 1;12(3):20. doi: 10.1167/tvst.12.3.20.

Abstract

PURPOSE

Accurate mapping of phosphene locations from visual prostheses is vital to encode spatial information. This process may involve the subject pointing to evoked phosphene locations with their finger. Here, we demonstrate phosphene mapping for a retinal implant using eye movements and compare it with retinotopic electrode positions and previous results using conventional finger-based mapping.

METHODS

Three suprachoroidal retinal implant recipients (NCT03406416) indicated the spatial position of phosphenes. Electrodes were stimulated individually, and the subjects moved their finger (finger based) or their eyes (gaze based) to the perceived phosphene location. The distortion of the measured phosphene locations from the expected locations (retinotopic electrode locations) was characterized with Procrustes analysis.

RESULTS

The finger-based phosphene locations were compressed spatially relative to the expected locations all three subjects, but preserved the general retinotopic arrangement (scale factors ranged from 0.37 to 0.83). In two subjects, the gaze-based phosphene locations were similar to the expected locations (scale factors of 0.72 and 0.99). For the third subject, there was no apparent relationship between gaze-based phosphene locations and electrode locations (scale factor of 0.07).

CONCLUSIONS

Gaze-based phosphene mapping was achievable in two of three tested retinal prosthesis subjects and their derived phosphene maps correlated well with the retinotopic electrode layout. A third subject could not produce a coherent gaze-based phosphene map, but this may have revealed that their phosphenes were indistinct spatially.

TRANSLATIONAL RELEVANCE

Gaze-based phosphene mapping is a viable alternative to conventional finger-based mapping, but may not be suitable for all subjects.

摘要

目的

准确绘制视觉假体的光点位置对于编码空间信息至关重要。这个过程可能涉及到受试者用手指指向诱发的光点位置。在这里,我们使用眼球运动来演示视网膜植入物的光点映射,并将其与基于传统的基于手指的映射的视网膜电极位置和以前的结果进行比较。

方法

三名脉络膜上视网膜植入物接受者(NCT03406416)指示光点的空间位置。单独刺激电极,受试者用手指(基于手指)或眼睛(基于凝视)移动到感知到的光点位置。使用 Procrustes 分析来描述测量的光点位置与预期位置(视网膜电极位置)的扭曲。

结果

三个受试者的基于手指的光点位置相对于预期位置在空间上都受到压缩,但保留了大致的视网膜排列(尺度因子范围为 0.37 至 0.83)。在两名受试者中,基于凝视的光点位置与预期位置相似(尺度因子分别为 0.72 和 0.99)。对于第三个受试者,基于凝视的光点位置与电极位置之间没有明显的关系(尺度因子为 0.07)。

结论

在三个测试的视网膜假体受试者中的两个中,可以实现基于凝视的光点映射,并且他们的衍生光点图与视网膜电极布局很好地相关。第三个受试者无法产生一致的基于凝视的光点图,但这可能表明他们的光点在空间上无法区分。

翻译相关性

基于凝视的光点映射是传统基于手指的映射的一种可行替代方法,但可能不适合所有受试者。