Influence of Operator Experience on Scanning Time and Accuracy with Two Different Intraoral Scanners - A Prospective Clinical Trial.

作者信息

Thomas Anjali Anna, Jain Ravindra Kumar

机构信息

Clinic of Orthodontics and Dentofacial Orthopaedics, Saveetha Dental College and Hospital, Saveetha Institute of Medical and Technical Sciences, Saveetha University, Tamil Nadu, India.

出版信息

Turk J Orthod. 2023 Mar 21;36(1):10-14. doi: 10.4274/TurkJOrthod.2022.2021.0220.

Abstract

OBJECTIVE

Operator experience and scanner type may influence the time taken and obtained accuracy of intraoral scanning. This study aimed to evaluate the influence of operator experience on the scanning time and correlate the accuracy of the scans taken with two different intraoral scanners (TRIOS 3, 3Shape and i500, Medit).

METHODS

In this trial, a total of 20 subjects who required intraoral scanning for orthodontic treatment were included. Intraoral scanning was done with two different scanners, TRIOS 3 and i500. One operator each with high (group 1), medium (group 2) and low (group 3) levels of experience performed intra-oral scanning with two different intraoral scanners. A One-Way ANOVA test was performed to assess the intergroup difference in scanning time and Kendall's tau's correlation test to determine the correlation between the experience of the operator and accuracy among the three groups using the two scanners. Also Independent samples t-test were performed to assess the intragroup differences in scanning time with two different scanners.

RESULTS

The scanning time was influenced by the type of intraoral scanner and operator experience (p<0.05). No significant correlation between operator experience and scanning accuracy in the three groups was noted (p>0.05). Statistically significant intragroup differences in scanning time between the two scanners were noted (p<0.05).

CONCLUSION

Less experienced operators took more time to scan a subject. Accuracy of scanning among three groups using two scanners was not influenced by the experience of the operator. Scanning with i500 IOS took more time than TRIOS.

摘要

目的

操作人员的经验和扫描仪类型可能会影响口腔内扫描所需的时间和获得的准确性。本研究旨在评估操作人员经验对扫描时间的影响,并关联使用两种不同口腔内扫描仪(TRIOS 3,3Shape公司;i500,Medit公司)进行扫描的准确性。

方法

在本试验中,纳入了总共20名需要进行正畸治疗口腔内扫描的受试者。使用两种不同的扫描仪TRIOS 3和i500进行口腔内扫描。一名经验丰富(第1组)、中等(第2组)和经验不足(第3组)的操作人员分别使用两种不同的口腔内扫描仪进行口腔内扫描。进行单因素方差分析以评估扫描时间的组间差异,并进行肯德尔tau相关性检验,以确定使用两种扫描仪的三组操作人员经验与准确性之间的相关性。还进行独立样本t检验以评估使用两种不同扫描仪时扫描时间的组内差异。

结果

扫描时间受口腔内扫描仪类型和操作人员经验的影响(p<0.05)。三组中未发现操作人员经验与扫描准确性之间存在显著相关性(p>0.05)。两组扫描仪之间在扫描时间上存在统计学显著的组内差异(p<0.05)。

结论

经验较少的操作人员扫描受试者花费的时间更多。使用两种扫描仪的三组扫描准确性不受操作人员经验的影响。使用i500口腔内扫描仪扫描比TRIOS花费的时间更多。