使用Biodex System 4 Pro对身体活跃的成年人进行等速踝关节、膝关节和髋关节力量测试-重测信度

Test-Retest Reliability of Isokinetic Ankle, Knee and Hip Strength in Physically Active Adults Using Biodex System 4 Pro.

作者信息

Tuominen Juho, Leppänen Mari, Jarske Heidi, Pasanen Kati, Vasankari Tommi, Parkkari Jari

机构信息

Facualty of Medicine and Health Technology, Tampere University, 33014 Tampere, Finland.

Tampere Research Center of Sports Medicine, UKK Institute for Health Promotion Research, 33500 Tampere, Finland.

出版信息

Methods Protoc. 2023 Mar 9;6(2):26. doi: 10.3390/mps6020026.

Abstract

BACKGROUND

The isokinetic dynamometry is considered a gold standard in muscle strength testing. The reliability of lower limb isokinetic strength measurements has not been thoroughly evaluated.

OBJECTIVE

To examine the test-retest reliability of isokinetic ankle plantar and dorsiflexion, ankle inversion and eversion, knee extension and flexion and hip abduction and adduction strength in physically active adults using Biodex System 4 Pro.

METHODS

Peak torques (PTs) and average peak torques (APTs) of the dominant and nondominant lower limbs were tested twice in 19 physically active adults 7 to 14 days apart.

RESULTS

The intraclass correlation coefficients (ICC) values varied from excellent to moderate and coefficient of variation of typical error (CV) values were 6.6-19.5%. Change in the mean expressed as a percent varied from -3.1% to 9.6%. There was no difference in the reliability between PT and APT values. Dominant lower limb was more reliable in every case if there was difference between limbs.

CONCLUSION

Test-retest reliability of isokinetic ankle, knee and hip strength in physically active adults using Biodex System 4 is mostly good or excellent. However, the observed range of the random variation has to be noted when using it in scientific follow-up studies or evaluation of patient progress in clinical settings.

摘要

背景

等速测力法被认为是肌肉力量测试的金标准。下肢等速力量测量的可靠性尚未得到充分评估。

目的

使用Biodex System 4 Pro检查身体活跃的成年人中,等速踝关节跖屈和背屈、踝关节内翻和外翻、膝关节伸展和屈曲以及髋关节外展和内收力量的重测可靠性。

方法

对19名身体活跃的成年人的优势和非优势下肢的峰值扭矩(PTs)和平均峰值扭矩(APTs)进行了两次测试,间隔7至14天。

结果

组内相关系数(ICC)值从优秀到中等不等,典型误差变异系数(CV)值为6.6-19.5%。以百分比表示的平均值变化范围为-3.1%至9.6%。PT值和APT值之间的可靠性没有差异。如果肢体之间存在差异,在每种情况下优势下肢都更可靠。

结论

使用Biodex System 4对身体活跃的成年人进行等速踝关节、膝关节和髋关节力量的重测可靠性大多良好或优秀。然而,在科学随访研究或临床环境中评估患者进展时使用它时,必须注意观察到的随机变化范围。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/23bd/10037623/f58d24eb8d71/mps-06-00026-g001.jpg