van Tittelboom Vanessa, Alemdaroglu-Gürbüz Ipek, Hanssen Britta, Heyrman Lieve, Feys Hilde, Desloovere Kaat, Calders Patrick, Van den Broeck Christine
Department of Rehabilitation Sciences, Ghent University, Ghent, Belgium.
Department of Rehabilitation Sciences, Katholieke Universiteit Leuven, Leuven, Belgium.
Front Sports Act Living. 2022 Feb 24;4:817216. doi: 10.3389/fspor.2022.817216. eCollection 2022.
This study aimed to analyze the reliability of concentric isokinetic strength assessments (knee and hip) using the Biodex System 4 in healthy children and assess the association with functional strength tests (sit-to-stand [STS], lateral-step-up [LSU]).
19 children (6-12 years) were included. Knee and hip flexion and extension, and hip abduction and adduction were tested at 60 and 90°/s.
Relative and absolute reliability at 60°/s tended to show better results compared to those at 90°/s. Intra class correlations (ICCs) of knee flexion and extension at 60°/s were good (0.79-0.89). For hip flexion, extension, abduction and adduction at 60°/s ICCs were moderate to good (0.53-0.83). The smallest detectable change (SDC) values (expressed in %) were highly variable. The SDC% for knee flexion and extension and hip abduction at 60°/s were around 50%. Positive associations were found between hip extension and abduction isokinetic strength and the STS test.
Concentric isokinetic strength assessments in healthy children using the Biodex System 4 were found reliable for knee flexion and extension and hip abduction. Limited associations were found between concentric isokinetic strength tests and functional strength tests.
本研究旨在分析使用Biodex System 4对健康儿童进行同心等速肌力评估(膝关节和髋关节)的可靠性,并评估其与功能性力量测试(坐立试验[STS]、侧步上台阶试验[LSU])之间的关联。
纳入19名儿童(6至12岁)。在60°/秒和90°/秒的速度下测试膝关节和髋关节的屈伸,以及髋关节的外展和内收。
与90°/秒时相比,60°/秒时的相对和绝对可靠性往往显示出更好的结果。60°/秒时膝关节屈伸的组内相关系数(ICC)良好(0.79 - 0.89)。60°/秒时髋关节屈伸、外展和内收的ICC为中等至良好(0.53 - 0.83)。最小可检测变化(SDC)值(以百分比表示)变化很大。60°/秒时膝关节屈伸和髋关节外展的SDC%约为50%。髋关节伸展和外展等速肌力与坐立试验之间存在正相关。
发现使用Biodex System 4对健康儿童进行同心等速肌力评估对于膝关节屈伸和髋关节外展是可靠的。同心等速肌力测试与功能性力量测试之间的关联有限。