El Abed Rashid, Al Raeesi Dana, Alshehhi Aisha, Alkhatib Zuhair, Khamis Amar H, Jamal Mohamed, Kim Hyeon-Cheol
Endodontic Department, Hamdan Bin Mohammed College of Dental Medicine, Mohammed Bin Rashid University of Medicine and Health Sciences, Dubai 505055, United Arab Emirates.
Endodontic Department, Emirates Health Services Establishment, Dubai, United Arab Emirates.
Materials (Basel). 2023 Mar 11;16(6):2261. doi: 10.3390/ma16062261.
This study aimed to assess the effect of mechanical loading and heating on the cyclic fatigue and torsional fracture resistances of heat-treated nickel-titanium files after usage and autoclaving. Sixty files (One Curve) were tested for cyclic fatigue and torsional fracture resistances using customized devices. The files were divided into three groups according to the test conditions ( = 10); new (group-N), used for simulated canal shaping (group-U), and sterilized after use (group-S). For cyclic fatigue resistances, the files were freely rotated in a curved metal canal under body temperature; the time elapsed to fracture was recorded and the numbers of cycles to fracture (NCF) were calculated. For the torsional resistances, the file tip was fixed and rotated until the file fractured. The maximum torsional load and distortion angle were recorded. The toughness was calculated. Fracture fragments were examined with a scanning electron microscope. Data were analyzed using a one-way analysis of variance and Tukey's post hoc test at the significance level of 95%. Group-U showed significantly higher NCF than group-S ( < 0.05). However, there was no significant differences between groups-N and -S in the NCF ( > 0.05). Group-N showed a significantly bigger distortional angle and higher torsional toughness than groups-U and -S, but the ultimate torsional strength did not have significant difference between the groups. Under the limitation of this study, autoclave sterilization after single-usage did not improve the fracture resistance of heat-treated One Curve nickel-titanium files.
本研究旨在评估机械加载和加热对使用及高压灭菌后的热处理镍钛锉循环疲劳和抗扭转断裂性能的影响。使用定制装置对60支锉(One Curve)进行循环疲劳和抗扭转断裂性能测试。根据测试条件将锉分为三组(每组n = 10):新锉(N组)、用于模拟根管预备的使用过的锉(U组)以及使用后灭菌的锉(S组)。对于循环疲劳性能,锉在体温下于弯曲的金属根管中自由旋转;记录至断裂的时间并计算断裂循环次数(NCF)。对于抗扭转性能,将锉尖固定并旋转直至锉断裂。记录最大扭转载荷和变形角度。计算韧性。用扫描电子显微镜检查断裂碎片。数据采用单因素方差分析和Tukey事后检验进行分析,显著性水平为95%。U组的NCF显著高于S组(P < 0.05)。然而,N组和S组的NCF之间无显著差异(P > 0.05)。N组的变形角度显著大于U组和S组,且扭转韧性更高,但各组的极限抗扭强度无显著差异。在本研究的局限性下,单次使用后高压灭菌并未提高热处理的One Curve镍钛锉的抗断裂性能。