Department of Endodontology, Faculty of Dentistry, Ege University, İzmir, Turkey.
Microsc Res Tech. 2020 Jul;83(7):790-794. doi: 10.1002/jemt.23469. Epub 2020 Mar 14.
The aim of this study was to evaluate the cyclic fatigue resistance of Dia-X, WaveOne Gold and One Curve files in a water bath at intracanal temperature (35°C). Thirty-nine instruments Dia-X, WaveOne Gold, and One Curve systems (n = 13) were tested in an artificial canal with a curvature angle of 60° and a radius of 3 mm. A water bath setup at a temperature of 35°C was used to simulate the intracanal temperature and time to fracture (TTF) as seconds was recorded. The mean data were analyzed statistically using one-way ANOVA, and post hoc Tukey test (p = .05). The fractured surface of the instruments was examined with scanning electron microscopy (SEM) and chemical composition of the instruments were investigated with energy dispersive X-ray spectroscopy (EDS). Statistically significant differences were detected in TTF values of all the systems as follows: One Curve > WaveOne Gold > Dia-X (p < .05). One Curve instruments demonstrated the highest TTF values in all the tested instruments. The EDS microanalysis revealed similar NiTi composition of on the surface of One Curve, WaveOne Gold, and Dia-X instruments. The novel manufacturing process, including C-wire heat treatment and the variable cross-section of the One Curve files, could be the main factors affecting the fatigue life of the instruments.
本研究旨在评估在 35°C 根管内温度的水浴环境中,Dia-X、WaveOne Gold 和 One Curve 锉的循环疲劳性能。将 39 个 Dia-X、WaveOne Gold 和 One Curve 器械(n = 13)分别在具有 60°曲率角和 3mm 半径的人工根管中进行测试。使用温度为 35°C 的水浴装置模拟根管内温度和断裂时间(TTF),记录以秒为单位的 TTF。使用单向方差分析和事后 Tukey 检验(p =.05)对平均数据进行统计学分析。使用扫描电子显微镜(SEM)检查器械的断裂表面,并使用能量色散 X 射线光谱仪(EDS)研究器械的化学成分。所有系统的 TTF 值均存在统计学差异,具体如下:One Curve>WaveOne Gold>Dia-X(p<.05)。One Curve 器械在所有测试器械中表现出最高的 TTF 值。EDS 微分析显示 One Curve、WaveOne Gold 和 Dia-X 器械表面的 NiTi 成分相似。新型制造工艺,包括 C 型丝热处理和 One Curve 锉的变截面,可能是影响器械疲劳寿命的主要因素。