Scott N R, Chakraborty J, Marks V
J Chromatogr. 1986 Mar 7;375(2):321-9. doi: 10.1016/s0378-4347(00)83724-x.
Chromatographic separation of methylxanthine metabolites was achieved using a Hypersil octadecylsilane column with a simple concave gradient elution programme of 0-12.75% acetonitrile in 1% tetrahydrofuran, pH 4.8, and the eluted components were detected by monitoring their absorption at 280 nm. An extraction procedure involving the formation of an ion-pair complex was developed which gave significant improvements over previously described methods including a shorter chromatographic run of 20 min. A thorough comparison of this procedure with a more convenient alternative involving direct injection of diluted urine specimens showed that the latter analysis was adequate for the quantitation of the major urinary metabolites of caffeine and theophylline.
使用Hypersil十八烷基硅烷柱,通过在1%四氢呋喃中0 - 12.75%乙腈的简单凹形梯度洗脱程序(pH 4.8)实现甲基黄嘌呤代谢物的色谱分离,通过监测洗脱组分在280 nm处的吸光度进行检测。开发了一种涉及形成离子对络合物的萃取程序,与先前描述的方法相比有显著改进,包括色谱运行时间缩短至20分钟。将该程序与更方便的直接注射稀释尿样的替代方法进行了全面比较,结果表明后者的分析足以对咖啡因和茶碱的主要尿代谢物进行定量。