Allergan Aesthetics, an AbbVie Company, 4 Millennium Way, Branchburg, NJ, 08876, USA.
Hernia. 2023 Aug;27(4):987-997. doi: 10.1007/s10029-023-02769-0. Epub 2023 Apr 8.
This study compared the in vitro/benchtop and in vivo mechanical properties and host biologic response to ovine rumen-derived/polymer mesh hybrid OviTex™ with porcine-derived acellular dermal matrix Strattice™ Firm.
OviTex 2S Resorbable (OviTex 2S-R) and Strattice morphology were examined in vitro using histology and scanning electron microscopy; mechanical properties were assessed via tensile test; in vivo host biologic response and explant mechanics were evaluated in a rodent subcutaneous model. Separately, OviTex 1S Permanent (OviTex 1S-P) and Strattice were evaluated in a primate abdominal wall repair model.
OviTex 2S-R demonstrated layer separation, whereas Strattice retained its structural integrity and demonstrated higher maximum load than OviTex 2S-R out-of-package (124.8 ± 11.1 N/cm vs 37.9 ± 5.5 N/cm, p < 0.001), 24 h (55.7 ± 7.4 N/cm vs 5.6 ± 3.8 N/cm, p < 0.001), 48 h (45.3 ± 14.8 N/cm vs 2.8 ± 2.6 N/cm, p = 0.003), and 72 h (29.2 ± 10.5 N/cm vs 3.2 ± 3.1 N/cm, p = 0.006) following collagenase digestion. In rodents, inflammatory cell infiltration was observed between OviTex 2S-R layers, while Strattice induced a minimal inflammatory response. Strattice retained higher maximum load at 3 (46.3 ± 27.4 N/cm vs 9.5 ± 3.2 N/cm, p = 0.041) and 6 weeks (28.6 ± 14.1 N/cm vs 7.0 ± 3.0 N/cm, p = 0.029). In primates, OviTex 1S-P exhibited loss of composite mesh integrity whereas Strattice integrated into host tissue with minimal inflammation and retained higher maximum load at 1 month than OviTex 1S-P (66.8 ± 43.4 N/cm vs 9.6 ± 4.4 N/cm; p = 0.151).
Strattice retained greater mechanical strength as shown by lower susceptibility to collagenase degradation than OviTex 2S-R in vitro, as well as higher maximum load and improved host biologic response than OviTex 2S-R in rodents and OviTex 1S-P in primates.
本研究比较了绵羊瘤胃衍生/聚合物网混合 OviTex™与猪源去细胞真皮基质 Strattice™Firm 的体外/台式和体内机械性能和宿主生物反应。
使用组织学和扫描电子显微镜观察 OviTex 2S 可吸收(OviTex 2S-R)和 Strattice 的形态;通过拉伸试验评估机械性能;在啮齿动物皮下模型中评估体内宿主生物反应和植入物力学。另外,在灵长类动物腹壁修复模型中评估了 OviTex 1S 永久性(OviTex 1S-P)和 Strattice。
OviTex 2S-R 显示出分层,而 Strattice 保持其结构完整性,并在包外(124.8 ± 11.1 N/cm 对 37.9 ± 5.5 N/cm,p < 0.001)、24 小时(55.7 ± 7.4 N/cm 对 5.6 ± 3.8 N/cm,p < 0.001)、48 小时(45.3 ± 14.8 N/cm 对 2.8 ± 2.6 N/cm,p = 0.003)和 72 小时(29.2 ± 10.5 N/cm 对 3.2 ± 3.1 N/cm,p = 0.006)时胶原酶消化后,其最大负载更高。在啮齿动物中,OviTex 2S-R 层之间观察到炎症细胞浸润,而 Strattice 引起的炎症反应很小。Strattice 在 3 周(46.3 ± 27.4 N/cm 对 9.5 ± 3.2 N/cm,p = 0.041)和 6 周(28.6 ± 14.1 N/cm 对 7.0 ± 3.0 N/cm,p = 0.029)时保持更高的最大负载。在灵长类动物中,OviTex 1S-P 表现出复合网完整性丧失,而 Strattice 与宿主组织整合,炎症反应轻微,并在 1 个月时保持比 OviTex 1S-P 更高的最大负载(66.8 ± 43.4 N/cm 对 9.6 ± 4.4 N/cm;p = 0.151)。
Strattice 在体外表现出较低的胶原酶降解易感性,因此保持了比 OviTex 2S-R 更高的机械强度,并且在啮齿动物中的最大负载更高,在灵长类动物中的宿主生物反应更好。