J Sep Sci. 2023 Jul;46(13):e2300067. doi: 10.1002/jssc.202300067. Epub 2023 Apr 20.
The conventional reverse fill/flush flow modulation for comprehensive two-dimensional gas chromatography requires a bleed capillary column to be connected to the outlet of the modulator channel. The purpose of this capillary, which does not contain the stationary phase, is to provide pressure resistance to the modulator channel flow. In this way, the desired modulator volumetric flow can be achieved, and channel over-filling can be avoided. Normally, the length and the internal diameter of the bleed capillary are chosen so as to obtain the modulator flow that is close to the flow of the first separation column. Thus, for any chosen set of chromatographic conditions, the required dimensions of the bleed capillary can be completely different, making the two-dimensional gas chromatography method development tedious and generating additional costs in consumables and analyst time. In this work, a tuneable pressure source generating a suitable backpressure was used instead of the fixed bleed capillary which has the advantage of the possibility to freely adapt the pressure resistance and generate the required modulator channel flow for any conditions. This set-up has been evaluated and compared in terms of the impact on the modulation performance to the set-up involving a fixed bleed capillary demonstrating comparable performance.
传统的反吹/冲洗流量调制方法用于全二维气相色谱法时,需要将一根放空毛细管柱连接到调制器通道的出口。该放空毛细管中不包含固定相,其作用是为调制器通道流提供耐压。这样,就可以实现所需的调制器体积流量,并且可以避免通道过填充。通常,放空毛细管的长度和内径选择为使调制器流量接近第一分离柱的流量。因此,对于任何选定的色谱条件,所需的放空毛细管尺寸可能完全不同,这使得二维气相色谱方法的开发变得繁琐,并在消耗品和分析人员时间方面产生额外成本。在这项工作中,使用可调压力源产生合适的背压代替固定放空毛细管,这种方法的优点是可以自由适应耐压,并为任何条件生成所需的调制器通道流。已经评估了该设置,并与涉及固定放空毛细管的设置进行了比较,结果表明其调制性能相当。