Dosimetry for Radiation Therapy and Diagnostic Radiology, Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, D-38116, Germany.
Phys Med Biol. 2023 May 30;68(11). doi: 10.1088/1361-6560/acce1d.
. The purpose of this investigation is to establish the water calorimeter as the primary standard in PTB's ultra-high pulse dose rate (UHPDR) 20 MeV reference electron beams.. The calorimetric measurements were performed at the PTB research linac facility using the UHPDR reference electron beam setups that enable a dose per pulse between about 0.1 Gy and 6 Gy. The beam is monitored by an in-flange integrating current transformer. The correction factors required to determine the absorbed dose to water were evaluated using thermal and Monte Carlo simulations. Measurements were performed with different total doses delivered per pulse by modifying the instantaneous dose rate within a pulse and by changing the pulse length. The obtained temperature-time traces were compared to the simulated ones to validate the thermal simulations. In addition, absorbed-dose-to-water measurements obtained using the secondary standard alanine dosimeter system were compared to measurements performed with the primary standard.. The simulated and measured temperature-time traces were shown to be consistent, within combined uncertainties, with one another. Measurements with alanine dosimeters proved to be consistent within= 1 of the total combined uncertainty with the absorbed dose to water determined using the primary standard.. The total relative standard uncertainty of absorbed dose to water determined using the PTB water calorimeter primary standard in UHPDR electron beams was estimated to be less than 0.5%, and the combined correction factors were found to deviate from 1 by less than 1% for both PTB UHPDR 20 MeV reference electron beams. The water calorimeter is therefore considered to be an established primary standard for the higher energy UHPDR reference electron beams.
. 本研究旨在将水热量计确立为 PTB 超高脉冲剂量率(UHPDR)20 MeV 参考电子束的主要标准。. 在 PTB 研究直线加速器设施中进行了热量计测量,使用了能够实现脉冲剂量约为 0.1 Gy 至 6 Gy 的 UHPDR 参考电子束设置。束流由法兰内积分电流互感器监测。通过改变脉冲内瞬时剂量率和改变脉冲长度,评估了确定水吸收剂量所需的校正因子。使用不同的总脉冲剂量进行了测量。获得的温度-时间曲线与模拟曲线进行了比较,以验证热模拟。此外,使用次级标准丙氨酸剂量计系统进行的水吸收剂量测量与使用主要标准进行的测量进行了比较。. 模拟和测量的温度-时间曲线在合并不确定度内彼此一致。丙氨酸剂量计的测量结果与使用主要标准确定的水吸收剂量在= 1 内一致。. 使用 PTB 水热量计主要标准在 UHPDR 电子束中确定的水吸收剂量的总相对标准不确定度估计小于 0.5%,并且发现两个 PTB UHPDR 20 MeV 参考电子束的组合校正因子偏差均小于 1%。因此,水热量计被认为是高能 UHPDR 参考电子束的既定主要标准。