Hip sensor location for flash interstitial glucose monitoring system provides adequate clinical accuracy in non-diabetic hypoglycemic dogs.

机构信息

Department of Small Animal Clinical Sciences, Texas A&M School of Veterinary Medicine & Biomedical Sciences, College Station, TX.

Department of Veterinary Physiology and Pharmacology, Texas A&M School of Veterinary Medicine & Biomedical Sciences, College Station, TX.

出版信息

Am J Vet Res. 2023 Apr 24;84(6). doi: 10.2460/ajvr.23.02.0024. Print 2023 Jun 1.

Abstract

OBJECTIVE

Flash glucose monitoring systems (FGMS) are frequently used for interstitial glucose monitoring in dogs with diabetes mellitus and are typically placed between the scapulae. We aimed to evaluate the variability between glucose measurements from FGMS placed in 2 locations (between the scapulae and over the hip) in non-diabetic dogs during rapidly induced hypoglycemia.

ANIMALS

24 apparently healthy colony dogs that were subjects in a teaching laboratory.

PROCEDURES

Prospective interventional study. FGMS sensors (FreeStyle Libre 14-day system) were placed between the scapulae and over the hip of all dogs. Regular insulin was administered (0.3 u/kg IV) and subsequent hypoglycemia was corrected. Before insulin administration and every 10 minutes over 90 minutes, interstitial glucose was recorded from both locations, and blood glucose was measured with a point-of-care blood glucose monitor (AlphaTRAK 2).

RESULTS

There was a constant bias of 5.6 mg/dL (95% limits of agreement: -26.3 to 37.5 mg/dL) between locations, but the proportional bias was not apparent. There was a correlation between FGMS locations (r = 0.731, P = < .001). Sensor site B was clinically accurate with 100% of paired samples within Parkes error grid zones A (83%) and B (17%) but did not meet the criteria for analytical accuracy.

CLINICAL RELEVANCE

In this model of induced hypoglycemia in healthy dogs, variation between measurements from FGMS locations was unlikely to have affected the clinical outcome. Placement of FGMS over the hip may be an acceptable alternative to placement between the scapulae, but the utility in hyperglycemic dogs is unknown.

摘要

目的

瞬态血糖监测系统(FGMS)常用于监测糖尿病犬的间质血糖,通常置于肩胛骨之间。我们旨在评估在快速诱导低血糖期间,非糖尿病犬 FGMS 放置在 2 个位置(肩胛骨之间和臀部上方)时葡萄糖测量值的变异性。

动物

24 只作为教学实验室对象的明显健康的群体犬。

程序

前瞻性干预研究。在所有犬的肩胛骨之间和臀部上方放置 FGMS 传感器(FreeStyle Libre 14 天系统)。给予常规胰岛素(静脉注射 0.3 u/kg),并纠正随后的低血糖。在胰岛素给药前和 90 分钟内每 10 分钟,从两个位置记录间质血糖,并使用即时血糖仪(AlphaTRAK 2)测量血糖。

结果

两个位置之间存在 5.6mg/dL 的恒定偏差(95%置信区间:-26.3 至 37.5mg/dL),但无明显的比例偏差。FGMS 位置之间存在相关性(r=0.731,P<0.001)。传感器位置 B 在 100%配对样本位于 Parkes 误差网格区 A(83%)和 B(17%)内,具有临床准确性,但不符合分析准确性标准。

临床意义

在健康犬诱导性低血糖模型中,FGMS 位置之间的测量值变化不太可能影响临床结果。FGMS 放置在臀部上方可能是肩胛骨之间放置的可接受替代方法,但在高血糖犬中的效用尚不清楚。