The FreeStyle Libre (Abbott Laboratories) is a flash glucose monitoring system (FGMS) that measures interstitial glucose concentration (IG). The system is factory-calibrated, easy to use, inexpensive, and could be useful for monitoring diabetic cats.
To evaluate the analytical and clinical accuracy of the FGMS in cats and establish the lag-time between IG and blood glucose concentration (BG).
Twenty client-owned diabetic cats and 7 purpose-bred healthy cats.
Prospective study. Blood glucose concentration was measured using a portable glucose meter validated for use in cats that served as a reference method for IG, as measured by FGMS. In diabetic cats, data were collected for sensor wearing time with different methods of application and accuracy across glycemic ranges. Accuracy was determined by fulfillment of ISO15197:2013 criteria. In healthy cats, lag-time between IG and BG was established after IV administration of exogenous glucose.
Good agreement between IG and BG was obtained (r = .93). Analytical accuracy was not achieved, whereas clinical accuracy was demonstrated with 100% of the results in zones A + B of the Parkes consensus error grid analysis. In the immediate 30 minutes after an IV bolus of glucose, when BG was increasing rapidly (approximately 2%/min), IG increased slowly, resulting in a difference of as much as 579 mg/dL, and no positive correlation between BG and IG was found.
The FGMS did not fulfill ISO requirements but is sufficiently accurate for glucose monitoring in cats, while considering the lag between IG and BG during periods of rapid changes in BG.
FreeStyle Libre(雅培公司)是一种即时血糖监测系统(FGMS),可测量间质葡萄糖浓度(IG)。该系统出厂时已校准,使用方便,价格低廉,对于监测糖尿病猫可能很有用。
评估 FGMS 在猫中的分析和临床准确性,并确定 IG 与血糖浓度(BG)之间的滞后时间。
20 只患有糖尿病的宠物猫和 7 只专门饲养的健康猫。
前瞻性研究。使用经过验证可用于猫的便携式血糖仪测量血糖浓度,该血糖仪作为 FGMS 测量的 IG 参考方法。在糖尿病猫中,根据不同的应用方法和血糖范围的准确性收集传感器佩戴时间的数据。准确性通过满足 ISO15197:2013 标准来确定。在健康猫中,在静脉注射外源性葡萄糖后建立 IG 和 BG 之间的滞后时间。
IG 和 BG 之间获得了良好的一致性(r =.93)。分析准确性未达到,而临床准确性通过 Parkes 共识误差网格分析的 A+B 区 100%的结果得到证明。在静脉注射葡萄糖后 30 分钟内,当 BG 快速增加(约 2%/分钟)时,IG 增加缓慢,导致差异高达 579mg/dL,并且未发现 BG 和 IG 之间存在正相关关系。
FGMS 不符合 ISO 要求,但考虑到 BG 快速变化期间 IG 和 BG 之间的滞后时间,它对于猫的血糖监测足够准确。