Dorr R T, Salmon S E, Marsh M E, Robertone A
Cancer Drug Deliv. 1986 Spring;3(2):139-46. doi: 10.1089/cdd.1986.3.139.
A new microprocessor-controlled portable infusion pump, the Pancretec Provider IV 2000 TM was tested in vitro and in vivo in cancer patients. The Provider is a rotary peristaltic, battery-powered pump capable of flow rates of 0.2 to 83 ml/hour with delivery volumes up to 1999 ml. Two programming modes are available: intermittent infusion and continuous infusion. Bench tests showed the flow rate accuracy to be within 96% of the desired rate. Flow rate precision was similarly excellent at +/- 2%. Clinical studies were performed in 14 ambulatory patients receiving continuous infusion antineoplastic or analgesic drugs over 5-60 days as outpatients. The majority of infusions delivered fluoropyrimidines via indwelling central venous access ports. Two of the patients received long term (60 days) continuous infusion of analgesics for pain control. Flow rate accuracy with the pumps was within +/- 5% in 90% of the 27 infusion courses (244 patient-days of continuous infusion therapy). A significant therapy deviation (interruption 10% of the desired course) occurred in three instances. One related to a procedural error (incomplete cartridge insertion into the pump), two were caused by fluid leakage which interrupted pump function. Defect alarms (both visual and audible) operated in both instances. A KVO flow rate of 0.1 ml/hr was also found to be adequate to maintain catheter patency in peripheral veins over a 24-hour period in two normal volunteers. We conclude that the Provider pump is an accurate, reliable and state of the art infusion system with wide clinical applicability.
一种新型的微处理器控制的便携式输液泵——Pancretec Provider IV 2000 TM,在癌症患者中进行了体外和体内测试。该输液泵是一种旋转蠕动式、电池供电的泵,流速为0.2至83毫升/小时,输送量可达1999毫升。有两种编程模式:间歇输注和连续输注。台架测试表明,流速精度在所需流速的96%以内。流速精密度同样出色,为±2%。对14名门诊患者进行了临床研究,这些患者作为门诊病人接受了5至60天的连续输注抗肿瘤或镇痛药物。大多数输注通过留置中心静脉通路端口输送氟嘧啶。两名患者接受了长期(60天)连续输注镇痛药以控制疼痛。在27个输注疗程(244个患者日的连续输注治疗)中的90%,使用该泵的流速精度在±5%以内。有3例出现了显著的治疗偏差(中断达到所需疗程的10%)。1例与操作失误(药筒未完全插入泵中)有关,2例是由液体泄漏导致泵功能中断。在这两种情况下,缺陷警报(视觉和听觉)均起作用。在两名正常志愿者中还发现,0.1毫升/小时的KVO流速足以在24小时内维持外周静脉导管通畅。我们得出结论,Provider输液泵是一种准确、可靠且具有广泛临床适用性的先进输液系统。