Dorr R T, Trinca C E, Griffith K, Dombrowsky P L, Salmon S E
Cancer Treat Rep. 1979 Feb;63(2):211-3.
The delivery of continuous iv bleomycin and 5-fluorouracil to nine ambulatory cancer patients was studied using a novel disposable portable infusion pump, the Ar/Med Infusor. Although the chemotherapy was satisfactorily delivered, there were significant mechanical problems, including variable flow rates and substantial drug cartridge failure. Most therapy courses were finished within 10% of the projected termination time. Serious venous complications did not occur. While further exploration of various devices for outpatient continuous iv therapy is recommended, we believe that technical improvements are needed for the Ar/Med Infusor.
使用一种新型一次性便携式输液泵Ar/Med Infusor,对9名门诊癌症患者进行了持续静脉输注博来霉素和5-氟尿嘧啶的研究。尽管化疗药物输注情况令人满意,但存在严重的机械问题,包括流速不稳定和药物 cartridge 大量故障。大多数疗程在预计结束时间的10% 内完成。未发生严重的静脉并发症。虽然建议进一步探索用于门诊持续静脉治疗的各种设备,但我们认为Ar/Med Infusor需要进行技术改进。