Confirming the progression of keratoconus is of paramount relevance to providing the appropriate treatment. Real change should be considered consistent over time. It must be greater than the variability of the measurement of the device used to monitor the cornea. The present study aimed to assess the intraobserver repeatability and intersession reproducibility of a Scheimpflug camera in measuring corneal parameters in virgin keratoconus and intrastromal corneal ring segments (ICRS) implantation eyes to discriminate real change from measurement noise. Sixty keratoconus and 30 ICRS eyes were included. Corneal parameters were determined in three consecutive measurements and were repeated 2 weeks later. The precision within the same session for all parameters was better in the keratoconic eyes, with mean repeatability limits 33% narrower (range 13% to 55%) compared with ICRS eyes. Mean reproducibility limits were 16% narrower (range +48% to -45%) compared with ICRS eyes. The cutoff values to consider a real corneal shape change were lower for virgin keratoconic than for ICRS, except for the thinnest corneal thickness and Stage C (ABCD system), which were the opposite. Corneal tomography measurements in ICRS eyes showed worse accuracy than in virgin keratoconus, which should be taken into account by practitioners in patients' follow up.
确认圆锥角膜的进展对于提供适当的治疗至关重要。真正的变化应被视为随时间持续存在。它必须大于用于监测角膜的设备测量的变异性。本研究旨在评估在测量初发性圆锥角膜和基质内角膜环片(ICRS)植入眼的角膜参数时,眼前节分析系统在观察者内重复性和不同测量期的再现性,以区分真实变化与测量噪声。研究纳入了60只圆锥角膜眼和30只ICRS植入眼。连续进行三次测量来确定角膜参数,并在2周后重复测量。在同一测量期内,所有参数在圆锥角膜眼中的精度更高,与ICRS植入眼相比,平均重复性极限窄33%(范围为13%至55%)。平均再现性极限比ICRS植入眼窄16%(范围为+48%至-45%)。除了最薄角膜厚度和C期(ABCD系统)外,初发性圆锥角膜中考虑角膜形状真正变化的临界值低于ICRS植入眼,而最薄角膜厚度和C期(ABCD系统)情况相反。ICRS植入眼的角膜断层扫描测量显示准确性低于初发性圆锥角膜,从业者在患者随访中应考虑到这一点。