0.55T 下的身体成分分析:可行性和精度。

Body composition profiling at 0.55T: Feasibility and precision.

机构信息

Electrical and Computer Engineering, University of Southern California, Los Angeles, California, USA.

Siemens Medical Solutions USA, Los Angeles, California, USA.

出版信息

Magn Reson Med. 2023 Sep;90(3):1114-1120. doi: 10.1002/mrm.29682. Epub 2023 May 1.

Abstract

PURPOSE

Body composition MRI captures the distribution of fat and lean tissues throughout the body, and provides valuable biomarkers of obesity, metabolic disease, and muscle disorders, as well as risk assessment. Highly reproducible protocols have been developed for 1.5T and 3T MRI. The purpose of this work was to demonstrate the feasibility and test-retest repeatability of MRI body composition profiling on a 0.55T whole-body system.

METHODS

Healthy adult volunteers were scanned on a whole-body 0.55T MRI system using the integrated body RF coil. Experiments were performed to refine parameter settings such as TEs, resolution, flip angle, bandwidth, acceleration, and oversampling factors. The final protocol was evaluated using a test-retest study with subject removal and replacement in 10 adult volunteers (5 M/5F, age 25-60, body mass index 20-30).

RESULTS

Compared to 1.5T and 3T, the optimal flip angle at 0.55T was higher (15°), due to the shorter T1 times, and the optimal echo spacing was larger, due to smaller chemical shift between water and fat. Overall image quality was comparable to conventional field strengths, with no significant issues with fat/water swapping or inadequate SNR. Repeatability coefficient of visceral fat, subcutaneous fat, total thigh muscle volume, muscle fat infiltration, and liver fat were 11.8 cL (2.2%), 46.9 cL (1.9%), 14.6 cL (0.5%), 0.1 pp (2%), and 0.2 pp (5%), respectively (coefficient of variation in parenthesis).

CONCLUSIONS

We demonstrate that 0.55T body composition MRI is feasible and present optimized scan parameters. The resulting images provide satisfactory quality for automated post-processing and produce repeatable results.

摘要

目的

身体成分 MRI 可捕捉体内脂肪和瘦组织的分布情况,并提供肥胖、代谢疾病和肌肉疾病的有价值的生物标志物,以及风险评估。已经为 1.5T 和 3T MRI 开发了高度可重复的方案。本研究旨在证明在 0.55T 全身系统上进行 MRI 身体成分分析的可行性和测试-再测试可重复性。

方法

健康成年志愿者在全身 0.55T MRI 系统上使用集成的体部射频线圈进行扫描。进行了实验以优化参数设置,如 TE、分辨率、翻转角、带宽、加速和过采样因子。最终方案通过 10 名成年志愿者(5 名男性/5 名女性,年龄 25-60 岁,BMI 为 20-30)的测试-再测试研究进行评估,包括志愿者移除和替换。

结果

与 1.5T 和 3T 相比,0.55T 的最佳翻转角较高(15°),这是由于 T1 时间较短,最佳回波间隔较大,这是由于水和脂肪之间的化学位移较小。整体图像质量与常规场强相当,没有明显的脂肪/水交换或信噪比不足的问题。内脏脂肪、皮下脂肪、总大腿肌肉体积、肌肉脂肪浸润和肝脏脂肪的重复性系数分别为 11.8cL(2.2%)、46.9cL(1.9%)、14.6cL(0.5%)、0.1pp(2%)和 0.2pp(5%)(括号内为变异系数)。

结论

我们证明了 0.55T 身体成分 MRI 是可行的,并提出了优化的扫描参数。生成的图像为自动后处理提供了令人满意的质量,并产生了可重复的结果。