整合压力级真耳-耦合腔差值测量的有效性和可靠性。

Validity and reliability of integrated pressure level real-ear-to-coupler difference measurements.

机构信息

School of Communication Sciences and Disorders, Western University, London, Ontario, Canada.

Health and Rehabilitation Sciences Graduate Program, Western University, London, Ontario, Canada.

出版信息

Int J Audiol. 2024 Jun;63(6):401-410. doi: 10.1080/14992027.2023.2205009. Epub 2023 May 2.

Abstract

OBJECTIVES

(1) To validate the measurement of foam-tip real-ear-to-coupler differences (wRECD) using an integrated pressure level (IPL) method and (2) to compare the reliability of this method to SPL-based measurement of the wRECD.

DESIGN

SPL-based wRECD and the proposed IPL wRECD measurement were completed bilaterally. Test-retest reliability of IPL wRECD was determined with full re-insertion into the ear canal and compared to published SPL wRECD test-retest data.

STUDY SAMPLE

22 adults with normal hearing and middle ear status were recruited.

RESULTS

Differences between SPL-based wRECD and IPL wRECD measurements were within 1.51 dB on average below 5000 Hz. At and above 5000 Hz, IPL wRECD exceeded SPL wRECDs by 6.11 dB on average. The average test-retest difference for IPL wRECD across all assessed frequencies was 0.75 dB with the greatest improvements in reliability found below 750 Hz and above 3000 Hz.

CONCLUSIONS

IPL wRECD yielded improved estimates compared to SPL wRECD in high frequencies, where standing-wave interference is present. Independence from standing wave interference resulted in increased wRECD values above 4000 Hz using the IPL measurement paradigm. IPL wRECD is more reliable than SPL wRECD, does not require precise probe-microphone placement, and provides a wider valid wRECD bandwidth than SPL-based measurement.

摘要

目的

(1) 验证使用集成压力级 (IPL) 方法测量泡沫尖端真耳耦合差 (wRECD) 的有效性,(2) 比较该方法与基于声压级 (SPL) 的 wRECD 测量的可靠性。

设计

双侧完成 SPL 基 wRECD 和拟议的 IPL wRECD 测量。通过完全重新插入耳道确定 IPL wRECD 的测试-重测可靠性,并与已发表的 SPL wRECD 测试-重测数据进行比较。

研究样本

招募了 22 名听力正常和中耳状态正常的成年人。

结果

SPL 基 wRECD 和 IPL wRECD 测量值之间的差异在 5000Hz 以下平均为 1.51dB。在 5000Hz 及以上,IPL wRECD 平均比 SPL wRECD 高出 6.11dB。在所有评估频率下,IPL wRECD 的平均测试-重测差异为 0.75dB,在 750Hz 以下和 3000Hz 以上,可靠性的提高最为显著。

结论

在存在驻波干扰的高频下,IPL wRECD 比 SPL wRECD 产生了更好的估计值。由于不受驻波干扰的影响,使用 IPL 测量范式,wRECD 值在 4000Hz 以上增加。与 SPL 基测量相比,IPL wRECD 更可靠,不需要精确的探头-麦克风放置,并且提供比 SPL 基测量更宽的有效 wRECD 带宽。