Intra-and inter-observer variability of point of care ultrasound measurements to evaluate hemodynamic parameters in healthy volunteers.

作者信息

Blanca Deborah, Schwarz Esther C, Olgers Tycho Joan, Ter Avest Ewoud, Azizi Nasim, Bouma Hjalmar R, Ter Maaten Jan Cornelis

机构信息

Department of Internal Medicine, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.

Department of Internal Medicine, Ospedale Maggiore Policlinico, Università Degli Studi di Milano, Milan, Italy.

出版信息

Ultrasound J. 2023 May 5;15(1):22. doi: 10.1186/s13089-023-00322-9.

Abstract

BACKGROUND

Point-of-care ultrasound (POCUS) is a valuable tool for assessing the hemodynamic status of acute patients. Even though POCUS often uses a qualitative approach, quantitative measurements have potential advantages in evaluating hemodynamic status. Several quantitative ultrasound parameters can be used to assess the hemodynamic status and cardiac function. However, only limited data on the feasibility and reliability of the quantitative hemodynamic measurements in the point-of-care setting are available. This study investigated the intra- and inter-observer variability of PoCUS measurements of quantitative hemodynamic parameters in healthy volunteers.

METHODS

In this prospective observational study, three sonographers performed three repeated measurements of eight different hemodynamic parameters in healthy subjects. An expert panel of two experienced sonographers evaluated the images' quality. The repeatability (intra-observer variability) was determined by calculating the coefficient of variation (CV) between the separate measurements for each observer. The reproducibility (inter-observer variability) was assessed by determining the intra-class correlation coefficient (ICC).

RESULTS

32 subjects were included in this study, on whom, in total, 1502 images were obtained for analysis. All parameters were in a normal physiological range. Stroke volume (SV), cardiac output (CO), and inferior vena cava diameter (IVC-D) showed high repeatability (CV under 10%) and substantial reproducibility (ICC 0.61-0.80). The other parameters had only moderate repeatability and reproducibility.

CONCLUSIONS

We demonstrated good inter-observer reproducibility and good intra-observer repeatability for CO, SV and IVC-D taken in healthy subjects by emergency care physicians.

摘要

背景

床旁超声(POCUS)是评估急性患者血流动力学状态的重要工具。尽管POCUS通常采用定性方法,但定量测量在评估血流动力学状态方面具有潜在优势。几种定量超声参数可用于评估血流动力学状态和心脏功能。然而,关于床旁环境下定量血流动力学测量的可行性和可靠性的数据有限。本研究调查了健康志愿者中POCUS定量血流动力学参数测量的观察者内和观察者间变异性。

方法

在这项前瞻性观察研究中,三名超声检查人员对健康受试者的八个不同血流动力学参数进行了三次重复测量。由两名经验丰富的超声检查人员组成的专家小组评估图像质量。通过计算每个观察者单独测量值之间的变异系数(CV)来确定重复性(观察者内变异性)。通过确定组内相关系数(ICC)来评估再现性(观察者间变异性)。

结果

本研究纳入了32名受试者,共获得1502张图像用于分析。所有参数均在正常生理范围内。每搏输出量(SV)、心输出量(CO)和下腔静脉直径(IVC-D)显示出高重复性(CV低于10%)和高再现性(ICC为0.61-0.80)。其他参数仅具有中等重复性和再现性。

结论

我们证明了急诊医生对健康受试者进行的CO、SV和IVC-D测量具有良好的观察者间再现性和良好的观察者内重复性。