Department of Applied Prosthodontics, Graduate School of Biomedical Sciences, Nagasaki University, Nagasaki, Japan.
J Prosthodont Res. 2024 Jan 16;68(1):40-49. doi: 10.2186/jpr.JPR_D_23_00032. Epub 2023 May 20.
This systematic review aimed to evaluate the effects of implant placement sites on the biomechanical behavior of implant-assisted removable partial dentures (IARPDs) using finite element analysis (FEA).
Two reviewers independently conducted manual searches of the PubMed, Scopus, and ProQuest databases for articles investigating implant location in IARPDs using FEA, according to the 2020 Systematic Reviews and Meta-analyses statement. Studies published in English up to August 1, 2022, were included in the analysis based on the critical question.
Seven articles meeting the inclusion criteria were systematically reviewed. Six studies investigated mandibular Kennedy Class I and one study investigated mandibular Kennedy Class II. Implant placement reduced the displacement and stress distribution of the IARPD components, including dental implants and abutment teeth, regardless of the Kennedy Class type and dental implant placement site. Most of the included studies showed that, based on the biomechanical behavior, the molar region, rather than the premolar region, is the preferred implant placement site. None of the selected studies investigated the maxillary Kennedy Class I and II.
Based on the FEA regarding mandibular IARPDs, we concluded that implant placement in both the premolar and molar regions improves the biomechanical behaviors of IARPD components, regardless of the Kennedy Class. Implant placement in the molar region results in more suitable biomechanical behaviors compared with implant placement in the premolar region in Kennedy Class I. No conclusion was reached for Kennedy Class II due to the lack of relevant studies.
本系统评价旨在通过有限元分析(FEA)评估种植体植入部位对种植体辅助可摘局部义齿(IARPD)生物力学行为的影响。
两名审查员根据 2020 年系统评价和荟萃分析声明,独立手动搜索 PubMed、Scopus 和 ProQuest 数据库,以查找使用 FEA 研究 IARPD 中种植体位置的文章。根据关键问题,分析中纳入了截至 2022 年 8 月 1 日发表的英文文章。
有 7 篇符合纳入标准的文章进行了系统回顾。6 项研究调查了下颌 Kennedy 类 I,1 项研究调查了下颌 Kennedy 类 II。种植体的植入减少了 IARPD 组件(包括种植体和基牙)的位移和应力分布,无论 Kennedy 类类型和种植体植入部位如何。大多数纳入的研究表明,基于生物力学行为,磨牙区而不是前磨牙区是首选的种植体植入部位。所选研究均未调查上颌 Kennedy 类 I 和 II。
根据针对下颌 IARPD 的 FEA,我们得出结论,无论 Kennedy 类如何,在前磨牙和磨牙区域植入种植体都可以改善 IARPD 组件的生物力学行为。与 Kennedy 类 I 中的前磨牙区相比,磨牙区的种植体植入可导致更合适的生物力学行为。由于缺乏相关研究,因此对于 Kennedy 类 II 没有得出结论。