To compare the quantitative parameters and qualitative image quality of dual-energy CT angiography (CTA) between two rapid kVp-switching dual-energy CT scanners.
Between May 2021 and March 2022, 79 participants underwent whole-body CTA using either Discovery CT750 HD (Group A, n = 38) or Revolution CT Apex (Group B, n = 41). All data were reconstructed at 40-keV and with adaptive statistical iterative reconstruction-Veo of 40%. The two groups were compared in terms of CT numbers of the thoracic and abdominal aorta, and the iliac artery, background noise, signal-to-noise ratio (SNR) of the artery, CT dose-index volume (CTDI), and qualitative scores for image noise, sharpness, diagnostic acceptability, and arterial depictions.
The median CT number of the abdominal aorta (p = 0.04) and SNR of the thoracic aorta (p = 0.02) were higher in Group B than in Group A, while no difference was observed in the other CT numbers and SNRs of the artery (p = 0.09-0.23). The background noises at the thoracic (p = 0.11), abdominal (p = 0.85), and pelvic (p = 0.85) regions were comparable between the two groups. CTDI was lower in Group B than in Group A (p = 0.006). All qualitative scores were higher in Group B than in Group A (p < 0.001-0.04). The arterial depictions were nearly identical in both two groups (p = 0.005-1.0).
In dual-energy CTA at 40-keV, Revolution CT Apex improved qualitative image quality and reduced radiation dose.
比较两种快速 kVp 切换双能 CT 扫描仪的双能 CT 血管造影(CTA)的定量参数和定性图像质量。
在 2021 年 5 月至 2022 年 3 月期间,79 名参与者使用 Discovery CT750 HD(A 组,n=38)或 Revolution CT Apex(B 组,n=41)进行全身 CTA。所有数据均以 40keV 重建,并使用 40%的自适应统计迭代重建-Veo。比较两组胸主动脉和腹主动脉及髂动脉的 CT 值、背景噪声、信噪比(SNR)、CT 剂量指数体积(CTDI)以及图像噪声、清晰度、诊断可接受性和动脉显示的定性评分。
B 组腹主动脉(p=0.04)和胸主动脉(p=0.02)的 CT 值中位数高于 A 组,而其他动脉 CT 值和 SNR 无差异(p=0.09-0.23)。两组胸(p=0.11)、腹(p=0.85)和骨盆(p=0.85)区域的背景噪声相似。B 组的 CTDI 低于 A 组(p=0.006)。B 组的所有定性评分均高于 A 组(p<0.001-0.04)。两组的动脉显示几乎相同(p=0.005-1.0)。
在 40keV 双能 CTA 中,Revolution CT Apex 可改善定性图像质量并降低辐射剂量。