Kovach J C, Werner F W, Palmer A K, Greenkey S, Murphy D J
J Hand Surg Am. 1986 Jul;11(4):562-6. doi: 10.1016/s0363-5023(86)80200-3.
Although numerous fixation techniques have been developed for performing small joint arthrodesis, no previous study of the biomechanical properties of these constructs has been published. The strength of specimens of arthrodesis of the proximal interphalangeal joint performed with cadaver material was determined for three-point anteroposterior bending, axial torsion, and lateral cantilever bending stress. Crossed Kirschner wires, intraosseous wiring, and figure of eight tension-band wiring were studied. Figure of eight tension-band wiring demonstrates superior strength in anteroposterior bending and torsion. For lateral bending stress, no significant difference exists among the techniques studied.