Yang Jihyun, Kim Jeehyun, Chun Byung-Chul, Lee Jae-Myeong
Division of Nephrology, Department of Internal Medicine, Sungkyunkwan University School of Medicine, Kangbuk Samsung Hospital, 29 Saemunan-ro, Jongno-gu, Seoul 03181, Republic of Korea.
Department of Preventive Medicine, Korea University College of Medicine, Goryeodae-ro 73, Seongbuk-gu, Seoul 02841, Republic of Korea.
Life (Basel). 2023 Apr 30;13(5):1119. doi: 10.3390/life13051119.
Bioelectrical impedance analysis (BIA) is gaining popularity as a tool for body composition assessment. Although BIA has been studied and validated in different populations, age groups, and clinical settings, including critically ill patients, there are concerns about BIA reproducibility and reliability for different device types and postures. This study aimed to evaluate the reliability of BIA using different devices, postures, and lead types. Cross-sectional observational data were collected from 74 healthy volunteers (32 women, 42 men). We used two types of devices, three types of postures (standing, sitting, and lying), and two lead types (clamp lead and adhesive lead) to measure the whole-body phase angle (phA) at a single frequency of 50 kHz. The measurements were validated using the intraclass correlation coefficient (ICC) and Bland-Altman plot analysis. All phA measurements recorded using the two types of devices, three different postures, and two types of leads were equivalent (mean ICC = 0.9932, 95% confidence interval (CI) 0.9905-0.0053, < 0.001). The average mean difference in phA was 0.31 (95% CI 0.16-0.46). The largest phA value was measured using BWA with an adhesive-type lead in the supine position. There were no differences between the standing and sitting positions. We compared the consistency and reliability of phA using two devices, two lead types, and three postures. Seven different phA were interchangeable in healthy volunteers.
生物电阻抗分析(BIA)作为一种身体成分评估工具正越来越受欢迎。尽管BIA已在不同人群、年龄组和临床环境(包括重症患者)中进行了研究和验证,但对于不同设备类型和姿势下BIA的可重复性和可靠性仍存在担忧。本研究旨在评估使用不同设备、姿势和导联类型时BIA的可靠性。从74名健康志愿者(32名女性,42名男性)收集横断面观察数据。我们使用两种类型的设备、三种姿势(站立、坐着和躺着)以及两种导联类型(夹式导联和粘贴式导联)在50kHz的单一频率下测量全身相位角(phA)。使用组内相关系数(ICC)和Bland-Altman图分析对测量结果进行验证。使用两种类型的设备、三种不同姿势和两种类型的导联记录的所有phA测量结果均等效(平均ICC = 0.9932,95%置信区间(CI)0.9905 - 0.0053,< 0.001)。phA的平均平均差异为0.31(95%CI 0.16 - 0.46)。最大的phA值是在仰卧位使用粘贴式导联的BWA测量的。站立和坐姿之间没有差异。我们比较了使用两种设备、两种导联类型和三种姿势时phA的一致性和可靠性。在健康志愿者中,七种不同的phA是可互换的。