Department of Mechanical Engineering, Ajou University, Suwon, Korea.
Department of Orthopaedic Surgery, Eulji university hospital, Daejeon, Korea.
BMC Musculoskelet Disord. 2023 Jun 7;24(1):465. doi: 10.1186/s12891-023-06579-4.
This study aimed to analyze the differences in the stability of fractures, stress distribution around the distal-most screw according to the length of the plate and the trajectory of the bolt in Pauwels type III femoral neck fracture using the femoral neck system (FNS).
Finite element models of Pauwels type III femoral neck fractures were established with surgical variations in the trajectory of the bolt (central, inferior, valgus, and varus) and length of the lateral plate (1- and 2-hole plate). The models were subsequently subjected to normal walking and stair-climbing loads.
The screw-holding cortical bone in subtrochanter in the model with a 2-hole plate and the bolt in the inferior trajectory and the models with 1-hole or 2-hole plate and the bolt in valgus trajectory had shown greater maximum principal strain than the models with central or varus trajectories. The gap and sliding distance on the fracture surface were larger with inferior or varus trajectories of the bolt and smaller with the valgus trajectory of the bolt under both loads, compared to those of the central trajectory.
For the fixation of Pauwels type III femoral neck fracture, the trajectory of the FNS bolt and the length of the plate affect the mechanical stability of the fracture and the strain of cortical bone around the distal-most screw. The surgical target should stay on the central trajectory of the bolt and the 2-hole plate's mechanical benefits did not exceed the risk.
本研究旨在分析使用股骨颈系统(FNS)时,根据钢板长度和螺栓轨迹,Pauwels 型 III 股骨颈骨折中骨折稳定性和最远端螺钉周围应力分布的差异。
建立了 Pauwels 型 III 股骨颈骨折的有限元模型,其中包括螺栓轨迹(中心、下、外翻和内翻)和外侧钢板长度(1 孔和 2 孔钢板)的手术变化。随后对模型进行了正常行走和爬楼梯负荷的测试。
在 2 孔钢板和下侧螺栓轨迹的模型中,以及在 1 孔或 2 孔钢板和螺栓外展轨迹的模型中,持螺钉皮质骨的最大主应变大于中心或内翻轨迹的模型。在两种负荷下,与中心轨迹相比,螺栓的下侧或内翻轨迹导致骨折表面的间隙和滑动距离更大,而螺栓的外展轨迹则更小。
对于 Pauwels 型 III 股骨颈骨折的固定,FNS 螺栓的轨迹和钢板的长度会影响骨折的力学稳定性和最远端螺钉周围皮质骨的应变。手术靶点应保持在螺栓的中心轨迹上,并且 2 孔钢板的力学优势不会超过风险。